Keithley Instruments, Inc.
Cleveland, Ohio 44139
DRN.
CKD.
APP.
DATE
DATE
DATE
PART NUMBER
LTR
REVISIONS
APP. DATE
FORM 28777A-SBG
BRUNING 40-21 62198-SBG
SPECIFICATIONS
SPEC-2440
A 25977
Released
SZ 5/22/01
B 27103
Revised
SZ 5/20/02
C 28003
Revised
SZ 3/17/03
HW 5/22/01
SZ 5/22/01
HW 3/26/03
Rev. C
Page 2 of 3
2440 and 2440-C SourceMeter
®
Specifications
MEASURE SPECIFICATIONS
1,2
VOLTAGE MEASUREMENT ACCURACY (LOCAL OR REMOTE SENSE)
RANGE
DEFAULT
RESOLUTION INPUT
RESISTANCE
ACCURACY (1 Year)
23°C
±
5°C
±
(% rdg. + volts)
200.00 mV
1 µV
>10 G
Ω
0.012% + 300 µV
2.00000 V
10 µV
>10 G
Ω
0.012% + 300 µV
10.0000 V
100 µV
>10 G
Ω
0.015% + 750 µV
40.0000 V
1 mV
>10 G
Ω
0.015% + 3 mV
TEMPERATURE COEFFICIENT (0°-18°C & 28°-50°C):
±
(0.15 × accuracy specification)/
°C.
CURRENT MEASUREMENT ACCURACY (LOCAL OR REMOTE SENSE)
RANGE
DEFAULT
RESOLUTION
VOLTAGE BURDEN
3
ACCURACY (1 Year)
23°C
±
5°C
±
(% rdg. + amps)
10.0000 µA
100 pA
<1 mV
0.027% + 700 pA
100.000 µA
1 nA
<1 mV
0.025% + 6 nA
1.00000 mA
10 nA
<1 mV
0.027% + 60 nA
10.0000 mA
100 nA
<1 mV
0.035% + 600 nA
100.000 mA
1 µA
<1 mV
0.055% + 6 µA
1.00000 A
10 µA
<1 mV
0.060% + 570 µA
5.00000 A
10 µA
<1 mV
0.10% + 3.42 mA
TEMPERATURE COEFFICIENT (0°-18°C & 28°-50°C):
±
(0.10 × accuracy specification)/
°C.
CONTACT CHECK
SPECIFICATIONS
SPEED:
350µs for verification and notification.
CONTACT CHECK:
2
Ω
15
Ω
50
Ω
No contact check
failure
<1.00
Ω
<13.5
Ω
<47.5
Ω
Always contact check
failure >3.00
Ω
>16.5
Ω
>52.5
Ω
RESISTANCE MEASUREMENT ACCURACY (LOCAL OR REMOTE SENSE)
RANGE
DEFAULT RESOLUTION
DEFAULT TEST CURRENT
NORMAL ACCURACY (23°C
±
5°C)
1 YEAR,
±
(% rdg. + ohms)
<0.20000
Ω
4
- -
Source
I
ACC
+ Meas. V
ACC
2.00000
Ω
10
µ
Ω
1 A
0.17% + 0.0003
Ω
20.0000
Ω
100
µ
Ω
100 mA
0.10% + 0.003
Ω
200.000
Ω
1 m
Ω
10 mA
0.08% + 0.03
Ω
2.00000 k
Ω
10
m
Ω
1 mA
0.07% + 0.3
Ω
20.0000 k
Ω
100 m
Ω
100
µA
0.06% + 3
Ω
200.000 k
Ω
1
Ω
10
µA
0.07% + 30
Ω
2.00000 M
Ω
10
Ω
5
µA
0.11% + 300
Ω
20.0000 M
Ω
100
Ω
0.5
µA
0.11% + 1 k
Ω
>20.0000 M
Ω
4
-
-
Source I
ACC
+ Meas. V
ACC
TEMPERATURE COEFFICIENT (0°-18°C & 28°-50°C):
±
(0.15 × accuracy
specification)/
°C.
SOURCE I MODE, MANUAL OHMS:
Total uncertainty = I source accuracy
+ V measure accuracy (4-wire remote sense).
SOURCE V MODE, MANUAL OHMS:
Total uncertainty = V source
ac I measure accuracy (4-wire remote sense).
6-WIRE OHMS MODE:
Available using active ohms guard and guard sense
(except on 1A and 5A ranges). Max. Guard Output Current: 50mA. Accuracy
is load dependent. Refer to White Paper no. 2033 for calculation formula.
GUARD OUTPUT IMPEDANCE:
<0.1
Ω
in ohms mode.
1. Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications,
except 200mV, 1A, 5A ranges, add 0.05%. For 0.01 PLC, add 0.05% of range to offset
specifications, except 200mV, 1A, 5A ranges, add 0.5%.
2. Accuracies apply to 2- or 4-wire mode when properly zeroed.
3. 4-wire mode.
4. Manual ohms only.
Содержание SourceMeter 2400
Страница 10: ......
Страница 13: ...1 PerformanceVerification...
Страница 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Страница 29: ...2 Calibration...
Страница 51: ...3 RoutineMaintenance...
Страница 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Страница 55: ...4 Troubleshooting...
Страница 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Страница 69: ...5 Disassembly...
Страница 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Страница 77: ...6 ReplaceableParts...
Страница 91: ...A Specifications...
Страница 111: ...B CommandReference...
Страница 123: ...C CalibrationPrograms...
Страница 130: ...C 8 Calibration Programs...
Страница 134: ......
Страница 136: ......