Performance Veri
fi
cation
9.
Increase the electronic load to the maximum load test current value.
Channel
Maximum load test current
Channel 1
1.47 A
Channel 2
1.47 A
Channel 3
4.9 A
10.
Enter the voltmeter reading into the table for checking DC Voltage Load
Regulation under the maximum load column for the channel. (See Table 19
on page 19.)
11.
Calculate the three values: Ref – Min, Max – Ref, and (Max – Ref)/0.98.
Enter the values into the appropriate columns. (See Table 19 on page 19.)
12.
Enter the largest of the three values calculated in step 11 into the Largest
column.
This completes the check for one channel. If needed, return to step 1 to run
through the check for the next channel.