Speci
fi
cations
Table 1: Series 2200 speci
fi
cations (cont.)
Parameter
Description
Current resolution,
nominal
1 mA
Current setting
accuracy
±(0.1% + 5 mA) at 25 °C ±5 °C
Voltage limit,
nominal
0 to 30 V
Current
temperature
coef
fi
cient, typical
±(0.1% + 5 mA) per °C outside the 25 °C ±5 °C standard range
Meter, nominal
Internal readback capability. Voltage and current can be measured internally.
Current readback
resolution, nominal
1 mA
Current
readback accuracy
±(0.1% of r 5 mA) at 25 °C ±5 °C
Current readback
temperature
coef
fi
cient, typical
±(0.1% + 5 mA) per °C outside of the 25 °C ±5 °C standard range
Voltage
readback accuracy
±(0.03% of r 10 mV) at 25 °C ±5 °C standard range
Voltage readback
resolution, nominal
1 mV
Voltage readback
temperature
coef
fi
cient, typical
±(0.03% of r 10 mV) per °C outside the 25 °C ±5 °C standard range
Line regulation
– voltage
Change over the full rated line voltage range: ±(0.01% + 3 mV)
Line regulation
– current
Change over full rated line voltage range: ±(0.1% + 3 mA)
Load regulation
– voltage
Change from 0 to 100% of full scale current: ±(0.01% + 3 mV)
Load regulation
– current
Change from 10 to 90% of full scale voltage: ±(0.01% + 3 mA)
Voltage ripple,
20 MHz bandwidth
< 3 mV
RMS
and 20 mV
p-p
< 1.5 mV
RMS
and 12 mV
p-p
, typical
Voltage ripple,
7 MHz bandwidth
< 1 mV
RMS
and 3 mV
p-p
Current ripple,
20 MHz bandwidth
< 5 mA
RMS
Common Mode
Current Noise
< 1.5
μ
A
RMS
, 700 Hz bandwidth.
2
Series 2200 Multichannel Programmable DC Power Supplies Technical Reference