Principles of Operation
31
2
2
2
2
2
2
2
2
2
Make sure that the signals are wired appropriately. Refer to the
KUSB-3100 Getting Started Manual
for an example of connecting an
edge-to-edge measurement application.
Rate Generation
Use rate generation mode to generate a continuous pulse output
signal from Counter 0 Out; this mode is sometimes referred to as
continuous pulse output or pulse train output.
The pulse output operation is enabled whenever the Counter 0 Gate
signal is active (high level, low level, or software gate). While the
pulse output operation is enabled, the counter outputs a high-to-low
going pulse with a pulse width of 50% continuously. As soon as the
operation is disabled, rate generation stops.
The period of the output pulse is determined by the C/T clock source
(either internal or external) and the clock divider used. You can
generate an output signal from Counter 0 Out with a frequency of
15 Hz to 12 MHz.
To specify rate generation mode, use software to specify the
counter/timer mode as rate, the C/T clock source as either internal or
external, the clock divider (2 to 65536), and the active gate type
(high-level or software gate). Refer to
for more information
about gate types.
Make sure that the signals are wired appropriately. Refer to the
KUSB-3100 Getting Started Manual
for an example of connecting a rate
generation application.
Содержание KUSB-3100
Страница 10: ...Contents x...
Страница 14: ...About this Manual xiv...
Страница 15: ...1 1 Overview Features 2 Supported Software 3...
Страница 18: ...Chapter 1 4...
Страница 46: ...Chapter 2 32...
Страница 47: ...33 3 Supported Device Driver Capabilities...
Страница 86: ...Chapter 4 72...
Страница 87: ...73 5 Troubleshooting General Checklist 74 Service and Support 77...
Страница 100: ...Chapter 6 86...
Страница 101: ...87 A Specifications...
Страница 112: ...Appendix A 98...
Страница 113: ...99 B Connector Pin Assignments...
Страница 115: ...101 C Reading from or Writing to the Digital Registers...