Specifications
95
A
A
A
A
A
A
A
A
A
lists the specifications for the external A/D trigger on the
KUSB-3100 module.
Internal clock frequency
15 Hz to 12 MHz
External clock divider
2 to 65536
a. The resolution is 16-bits for rate generation mode.
Table 7: External Trigger Specifications
Feature
KUSB-3100 Specifications
Input type
Low-level or falling edge
sensitive
Logic family
TTL
Inputs
Input logic load:
High input voltage:
Low input voltage:
Low input current:
Level sensitive
1 TTL Load
2.4 V min
0.8 V max
–
0.4 mA max
Minimum pulse width
High:
Low:
200 ns
200 ns
Triggering modes
Single scan:
Continuous scan:
Yes
Yes
Table 6: C/T Subsystem Specifications (cont.)
Feature
KUSB-3100 Specifications