Section 2: Performance verification
Model DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
2-36
DMM6500-905-01 Rev. A/July 2018
Figure 18: Front panel 4-wire shorting plug orientation
8. Verify that the 1
Ω range is within specification (see the following table).
9. Repeat verification for the 10
Ω to 100 kΩ ranges.
Verify 4-wire resistance zero values
Range (Ω)
Lower limit
Upper limit
1
–2.00E-04
2.00E-04
10
–2.00E-04
2.00E-04
100
–2.00E-03
2.00E-03
1 k
–6.00E-03
6.00E-03
10 k
–6.00E-02
6.00E-02
100 k
–1.00E+00
1.00E+00
Verify DC voltage zero values using the 4-wire short
To verify DC voltage zero values:
1. Leave the short connected as described in
Verify resistance zero values using a 4-wire short
2. Press the
FUNCTION
key.
3. Select the
DC Voltage
function.
4. Press the
HOME
key.
5. Set the range to
1000 V
.
DC voltage verification is done in descending range order, starting with the 1000 V range and
finishing on the 100 mV range.
6. Verify that the 1000 V range zero is within specification. See the table below.
7. Verify that the 100 V to 100 mV range zero is within specification.