Section 8: Grading and binning resistors
DMM6500 6½ Digit Multimeter User's Manual
8-4
DMM6500-900-01Rev. A / April 2018
Trigger model template: grade and binning test
The trigger model template contains the settings for the number of components, digital I/O, and limits.
The command parameters for the template are described in the following command and table.
SCPI command usage:
:TRIGger:LOAD "GradeBinning", <components>, <startInLine>, <startDelay>,
<endDelay>, <limit1High>, <limit1Low>, <limit1Pattern>, <allPattern>,
<limit2High>, <limit2Low>, <limit2Pattern>, <limit3High>, <limit3Low>,
<limit3Pattern>, <limit4High>, <limit4Low>, <limit4Pattern>, "<bufferName>"
TSP command usage:
trigger.model.load("GradeBinning",
components
,
startInLine
,
startDelay
,
endDelay,
limit1High
,
limit1Low
,
limit1Pattern
,
allPattern
,
limit2High
,
limit2Low
,
limit2Pattern
,
limit3High
,
limit3Low
,
limit3Pattern
,
limit4High
,
limit4Low
,
limit4Pattern
,
bufferName
)
Parameter list
components
100
startInLine
Digital I/O line 5
startDelay
100 ms
endDelay
100 ms
limit1High
R = 100
Ω, P = 20%, 100+20% = 120 Ω
limit1Low
R = 100
Ω, P = 20%, 100-20% = 80 Ω
limit1Pattern
Bin 1 Fail Pattern 15: drive all digital I/O lines high (1111)
allPattern
All Pass Pattern 4: drive line 3 high (0100)
limit2High
R = 100
Ω, P =10%, 100+10% = 110 Ω
limit2Low
R = 100
Ω, P =10%, 100-10% = 90 Ω
limit2Pattern
Bin 2 Fail Pattern 1: drive line 1 high (0001)
limit3High
R = 100
Ω, P =5%, 100+5% = 105 Ω
limit3Low
R = 100
Ω, P =5%, 100-5% = 95 Ω
limit3Pattern
Bin 3 Fail Pattern 2: drive line 2 high (0010)
limit4High
R = 100
Ω, P =1%, 100+1% = 101 Ω
limit4Low
R = 100
Ω, P =1%, 100-1% = 99 Ω
limit4Pattern
Bin 4 Fail Pattern 3: drive line 1 and 2 high (0011)
bufferName
The reading buffer is set to
bufferVar