Models 6517 and 6517B Electrometer/High Resistance Meters
The Model 8009 Resistivity Fixture is designed to fully support the enhanced
resistivity measurement capability of the Model 6517 and Model 6517B
Electrometer/High Resistance Meters. Models 6517 and 6517B employ the
ASTM D-257 measurement method, and display measurements in resistance,
surface resistivity, or volume resistivity. All the Model 8009 electrode constants
are programmed into Models 6517 and 6517B. A built-in high voltage source
provides test voltages up to 1000 volts.
Models 6517 and 6517B offer special features for sophisticated, precise
measurement of resistivity. Both models can automatically implement a “bias-
measure” sequence in which the test voltage is applied for a programmed time to
permit resistivity to reach equilibrium, after which the measurement can be made
at some desired voltage. Models 6517 and 6517B can also measure and record
Fi
Mod
Figure 2-2
Model 8009 schematic diagram
6