5-12
6517B-900-01 Rev. A / Jun 2008
Section 5: Remote Operation
Model 6517B Electrometer User’s Manual
:PDTime <NRf>
Specify pre-discharge time; 0 to 99999.9s
:SVOLtage <NRf>
Specify bias voltage; -1000 to 1000
:STIMe <NRf>
Specify bias time; 0 to 99999.9
:MVOLtage <NRf>
Specify measure voltage; -1000 to 1000
:MTIMe <NRf>
Specify measure time; 0 to 99999.9s
:DTIMe <NRf>
Specify discharge time; 0 to 99999.9s
:VRESistivity
Volume resistivity test path:
:PDTime <NRf>
Specify pre-discharge time; 0 to 99999.9s
:SVOLtage <NRf>
Specify bias voltage; -1000 to 1000
:STIMe <NRf>
Specify bias time; 0 to 99999.9s
:MVOLtage <NRf>
Specify measure voltage; -1000 to 1000
:MTIMe <NRf>
Specify measure time; 0 to 99999.9s
:DTIMe <NRf>
Specify discharge time; 0 to 99999.9s
:SIResistance
Surface insulation resistance test path:
:SVOLtage <NRf>
Specify bias voltage; -1000 to 1000
:STIMe <NRf>
Specify bias time; 0 to 99999.9s
:MVOLtage <NRf>
Specify measure voltage; -1000 to 1000
:MTIMe <NRf>
Specify measure time; 0 to 99999.9s
:SQSWeep
Square wave sweep test path:
:HLEVel <NRf>
Specify high-level voltage; -1000 to 1000
:HTIMe <NRf>
Specify high-level time; 0 to 99999.9s
:LLEVel <NRf>
Specify low-level voltage; -1000 to 1000
:LTIMe <NRf>
Specify low-level time; 0 to 99999.9s
:COUNt <NRf>
Specify number cycles
:STSWeep
Staircase sweep test path:
:STARt <NRf>
Specify start voltage; -1000 to 1000
:STOP <NRf>
Specify stop voltage; -1000 to 1000
:STEP <NRf>
Specify step voltage; -1000 to 1000
:STIMe <NRf>
Specify step time; 0 to 99999.9s
Status register
The status register structure of the Model 6517B lets you monitor and act upon numerous events
that occur. Many programming decisions can be made by monitoring the measurement event
register and the operation event register. See Section 14 of the User’s Manual for details of the
status registers.
:STATus
STATus subsystem
:MEASurement
Path for measurement status registers:
[:EVENt]?
Use to read the measurement event register
:OPERation
Path for operation status registers:
[:EVENt]?
Use to read the operation event register
When an event register is read, a decimal value is sent to the computer. The binary equivalent
determines which bits in the register are set. For example, a decimal value of 33 indicates that bits
B0 and B5 are set. An event register is cleared when it is read. Sending
*CLS
clears all event
registers.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Содержание 6517B
Страница 3: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...
Страница 5: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...
Страница 89: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...
Страница 90: ...12 06 Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...