6517B-900-01 Rev. A / Jun 2008
Return to
5-11
Model 6517B Electrometer User’s Manual
Section 5: Remote Operation
Trigger model
Performing a system preset (
:SYSTem:PRESet
) configures the advanced trigger model for typical
instrument operation. After a system preset, many advanced operations can be accomplished
using the initiate commands,
:ABORt
, and the trigger (measure) layer commands.
:INITiate
[:IMMediate]
Take the instrument out of idle (ARM indicator on)
:CONTinuous <b>
Enable or disable continuous initiation
:ABORt
Reset trigger system; instrument stays in idle if continuous
initiation is disabled (off)
:TRIGger
Path to configure the advanced trigger layer:
:SOURce <name>
Select control source:
IMMediate, TIMer, MANual, BUS,
TLINk, EXTernal
or
HOLD
:TIMer <n>
Set timer interval (sec): 0.001 to 9999999.999
:DELay <n>
Set delay between measurements (sec) 0 to 999999.999
:SYSTem
:MACRo
:TRIGger
Path to configure basic triggering
[:EXECute]
Select basic triggering
:MODE <name>
Specify trigger mode (
CONTinuous
or
ONEShot
)
:SOURce <name>
Specify source (
IMMediate, MANual, BUS,
EXTernal
or
TIMer
)
:TIMer <n>
Specify timer interval; 0.001 to 99999.999 (sec.)
Test sequences
:TSEQuence
TSEQuence
subsystem
:ARM
Arms selected test sequence
:ABORt
Stops a test sequence in process
:TYPE <name>
Specify test;
DLEakage, CLEakage, CIResistance,
RVCoefficient, SRESistivity, VRESistivity,
SIResistivity, SQSWeep, STSWeep
:TSOurce <name>
Specify trigger source;
MANual, IMMediate, BUS,
TLINk, EXTernal, LCLosure
:TLINe <NRf>
Specify
TLINK
line; 1 to 6
DLEakage
Diode leakage test path:
:STARt <NRf>
Specify start voltage; -1000 to 1000
:STOP <NRf>
Specify stop voltage; -1000 to 1000
:STEP <NRf>
Specify step voltage; -1000 to 1000
:MDELay <NRf>
Specify measure delay; 0 to 99999.9s
:CLEakage
Capacitor leakage test path:
:SVOLtage <NRf>
Specify bias voltage (-1000 to 1000)
:STIMe <NRf>
Specify bias time; 0 to 99999.9s
:DTIMe <NRf>
Specify discharge time; 0 to 9999.9s
:CIResistance
Cable insulation resistance test path:
:SVOLtage <NRf>
Specify bias voltage; -1000 to 1000
:STIMe <NRf>
Specify bias time; 0 to 99999.9s
:RVCoefficient
Resistor voltage coefficient test path:
:SVOLtage[1] <NRf>
Specify source voltage 1; -1000 to 1000
:MDELay[1] <NRf>
Specify measure delay 1; 0 to 99999.0s
:SVOLtage2 <NRf>
Specify source voltage 2; -1000 to 1000
:MDELay2 <NRf>
Specify measure delay 2; 0 to 99999.9s
:SRESistivity
Surface resistivity test path:
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Содержание 6517B
Страница 3: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com ...