Series 3700 System Switch/Multimeter Quick Start Guide
Section 4
:
Frequently Asked Questions
Document Number: 3700S-903-01 Rev. A / August 2007
4-9
How do I create a scan (with reading buffer)?
NOTE
A reading buffer has to be created and selected before it can be used to store data.
To create a scan that sends it's measured readings to a reading buffer:
1. If a reading buffer that you wish to use has not already been created, create a reading
buffer:
Pressing the
STORE
key. The RD BUF ACTION MENU will be displayed.
Highlight CREATE and press the
ENTER
key.
Edit the reading buffer name as desired. Press the wheel (or the
ENTER
key) to enter edit
mode (the EDIT annunciator will light). Turn the wheel to change the characters values, use
the
CURSOR
keys to change the character being edited. Press the
ENTER
key to leave the
edit mode.
Enter the CAPACITY you would like the buffer to have in number of readings. Press the
ENTER
key to save the capacity.
2. Select the desired reading buffer:
Pressing the
STORE
key. The RD BUF ACTION MENU will be displayed. Highlight SELECT
and press the
ENTER
key. Highlight the desired buffer and press the
ENTER
key.
3. Add channels and DMM configurations.
Use the wheel to configure a channel that you would like added to the scan. Change to any of
the available channels and setup that channel with the associated DMM configuration that is
desired for the measurement.
The first press of the wheel (or
ENTER
key) will select the slot (from 1-6).
The second press of the wheel (or
ENTER
key) will select the channel number (card
dependent).
The third press of the wheel (or
ENTER
key) will select a range (card dependent).
The fourth press of the wheel (or
ENTER
key) will select the function (card dependent).
4. Press the
INSERT
key to add the step to the scan. Steps are added in the same order that
they have been entered from the front panel.
5. Continue adding steps as desired by repeating steps 3 and 4.
6. When finished adding steps to the scan, press the
SCAN
key. Select EXECUTE or
EXECUTE BACKGROUND to run the scan. Press the
TRIG
key to trigger measurements.
When finished, press the
REC
key and turn the wheel to view the contents of the reading
buffer.