Section 9: Constant current adjustment
Series 2260B Programmable Power Supplies Verification and Adjustment Manual
9-4
077104802 / November 2016
12. Press the voltage knob to set the half-scale current value (C124).
The "x" in the diagram above indicates that the value is not fixed.
13. Rotate the current knob to set the half-scale current value on the 2260B display so that it is the
same as the reading on the PCS-1000 current meter.
14. Press the voltage knob to enter the output current full-scale adjustment (C125).
15. Press the voltage knob to set the full-scale value (C126).
The "x" in the diagram above indicates that the value is not fixed.
1. Rotate the current knob to set the full-scale value on the 2260B display so that it is the same as
the reading on the PCS-1000 current meter.
2. Press the voltage knob to enter the OCP (overcurrent protection) adjustment (C127).
3. Press the voltage knob to start the automatic OCP adjustment (C128).