Series 2260B Programmable Power Supplies Verification and Adjustment Manual
Section 5: Verification test record forms
077104802 / November 2016
5-5
Constant current tests:
Current programming
and measurement
Model
Minimum
specifications
Results
Maximum specifications
Minimum current out
80-13
-10 mA
+10 mA
80-27
-30 mA
+30 mA
80-40
-40 mA
+40 mA
Measurement accuracy 80-13
-20 mA
+20 mA
80-27
-40 mA
+40 mA
80-40
-50 mA
+50 mA
Rated current out
80-13
12.87 A
13.13 A
80-27
26.943 A
27.057 A
80-40
39.92 A
40.080 A
Measurement accuracy 80-13
12.967 A
13.033 A
80-27
26.933 A
27.067 A
80-40
39.91 A
30.09 A
CC load regulation
Model
Minimum
specifications
Results
Maximum specifications
80-13
-18 mA
+18 mA
80-27
-32 mA
+32 mA
80-40
-45 mA
+45 mA
CC line regulation
Model
Minimum
specifications
Results
Maximum specifications
80-13
-18 mA
+18 mA
80-27
-32 mA
+32 mA
80-40
-45 mA
+45 mA