Series 2260B Programmable Power Supplies Verification and Adjustment Manual
Section 5: Verification test record forms
077104802 / November 2016
5-7
Constant current tests:
Current programming
and measurement
Model
Minimum
specifications
Results
Maximum specifications
Minimum current out
250-4
-5 mA
+5 mA
250-9
-10 mA
+10 mA
250-13
-15 mA
+15 mA
Measurement accuracy 250-4
-5 mA
+5 mA
250-9
-10 mA
+10 mA
250-13
-20 mA
+20 mA
Rated current out
250-4
3.991 A
4.001 A
250-9
8.9810 A
9.0190 A
250-13
12.972 A
13.028 A
Measurement accuracy 250-4
3.991 mA
4.001 mA
250-9
8.981 A
9.019 A
250-13
12.967 A
13.033 A
CC load regulation
Model
Minimum
specifications
Results
Maximum specifications
250-4
-9.000 mA
+9.000 mA
250-9
-14.000 mA
+14.000 mA
250-13
-18.00 mA
+18.00 mA
CC line regulation
Model
Minimum
specifications
Results
Maximum specifications
250-4
-9.5 mA
+9.5 mA
250-9
-14 mA
+14 mA
250-13
-18.5 mA
+18.5 mA