Series 2260B Programmable Power Supplies Verification and Adjustment Manual
Section 8: Constant voltage adjustment
077104802 / November 2016
8-3
5. Use the current knob to enter "constant voltage" adjustment mode. To enter adjustment mode,
rotate the current knob until "1" is displayed on the bottom display (C000:1).
6. Press the voltage knob to enter voltage adjustment (C110).
7. Press the voltage knob two times to enter the voltage offset adjustment (C111).
8. Press the voltage knob to set the offset voltage value (C112).
The "x" in the diagram above indicates that the value is not fixed.
9. Rotate the current knob to set the offset value on the 2260B display so that it is the same as the
reading on the PCS-1000 voltmeter.
10. Press the voltage knob to enter the output voltage middle adjustment (C113).
11. Press the voltage knob to set the middle voltage value.