Connect
Remote sense connection
You can use the remote sense connection for better
measurement precision.
To connect the DUT to the power supply using remote
sense:
1. Disconnect the wires or short clips from + to S+,
and – to S–.
2. Connect the S+ to the positive (+) terminal of the DUT,
and connect the S− to the negative (−) terminal of the
DUT, as shown in the following wiring diagram.
CAUTION
To ensure the stability of the system, use shielded
twisted pair cable between the remote sense terminals
and the instrument under test.
To avoid damage to the Series 2230 instrument, ensure
that the polarity of the connections is correct.