![background image](http://html2.mh-extra.com/html/keithley/2001-tcscan/2001-tcscan_instruction-manual_3982931013.webp)
1-1
1.1
Introduction
This section contains general information about the Model
2001-TCSCAN General Purpose/Thermocouple Scanner
Card, which is designed to be used with the Model 2000,
2001, 2002, and 2010 DMMs to make accurate multi-channel
thermocouple measurements. The DMMs will automatically
convert type J, K, T, E, R, S, and B thermocouple voltages to
Celsius, Fahrenheit, or Kelvin temperature readings.
The Model 2001-TCSCAN can also be used for a variety of
nine-channel, mixed-signal switching applications. The
500nV-125V signal voltage range of the Model 2001-
TCSCAN makes it well suited for both low and high signal
levels.
Section 1 is arranged in the following manner:
1.2
Features
1.4
Manual addenda
1.5
Safety symbols and terms
1.6
Specifications
1.7
Unpacking and inspection
1.8
Repacking for shipment
1.9
Optional accessories
1.2
Features
The Model 2001-TCSCAN is a scanner card designed to be
installed in the Model 2000, 2001, 2002, or 2010 Multimeter.
Key features include:
• Built-in reference junction (channel 1).
• Low contact potential and offset current.
• Input connectors are in contact with an isothermal block
to minimize temperature differences.
• Nine channels of 2-pole relay input.
• Four channel pairs configurable for 4-pole operation.
• Multiplex one of nine 2-pole, or one of four 4-pole
channels into the DMM.