background image

 

Page 7 

 

 

DOC1049R (22 February 2024 12:07) 

www.keit.co.uk 

Back to Table of Contents 

A peak at 1400 cm

-1

 indicates that there is some water on the probe. This could indicate poor 

purge quality. 

 

Figure 5: Peak indicating the water in the spectrum 

A  negative  peak  indicates that  although the  probe  is  clean  now,  it  was not clean  when the 
background  was  taken.  Another  background  must  be  taken  before  any  spectra  can  be 
measured. 

 

Figure 6: Negative peaks indicating the background was taken when the probe was unclean or not fully 
purged. 

1.4.  Taking a background 

FTIR requires the collection of a background spectrum (also called a reference or baseline 
scan)  before  any  absorption  scans  can  be  performed.  This  must  be  performed  before  the 
probe is inserted into the reaction/process of interest. The process for collecting a reference 
scan can be found in 

DOC0893 KeitSpec Software User Manual,

 Section 1, 

Getting Started

.

 

Содержание IRmadillo DOC1049R

Страница 1: ...DOC1049R 22 February 2024 12 07 DOC1049R IRmadilloNA User Manual Hardware Model Number ASM0627 09 N Cx O Gx D3x www keit co uk support keit co uk...

Страница 2: ...nd 7 2 Maintenance 8 2 1 Spectrum health 8 2 1 1 Use at high temperature 8 2 1 2 Emitter ageing 9 2 2 Remote health check 9 2 3 Cleaning the fibre optic data cable connectors 9 2 3 1 Cleaning the Sour...

Страница 3: ...e used The output from this purge gas generator should have a minimum quality according to ISO8573 1 2010 of Class 1 2 1 dust water oil or better Keit can recommend suitable generators that achieve th...

Страница 4: ...If the purge tubing is disconnected the blanking plugs must be fitted to the inlet connections on the probe and body to maintain the IP rating and exclude dust The IRmadilloNA requires a stabilisation...

Страница 5: ...reduced diameter section of the pipe The minimum flow velocity should be 1 5 m s 5 ft s 4 Install the IRmadilloNA so that the probe tip is fully in the flow We recommend putting the probe tip as close...

Страница 6: ...d The following are suitable Water Acetone Alcohols Surfactant solutions Alkanes hexane cyclohexane etc Ethers Once exposed to a suitable solvent wipe the ATR with a clean cloth The metal tube of the...

Страница 7: ...of interest within the spectrum is between 800 and 1800 cm 1 This section of the absorbance plot should maintain a relatively flat line with no consistently repeating peaks Cleaning Mode can be used t...

Страница 8: ...other background must be taken before any spectra can be measured Figure 6 Negative peaks indicating the background was taken when the probe was unclean or not fully purged 1 4 Taking a background FTI...

Страница 9: ...ment 2 1 1 Use at high temperature Figure 7 Probe is heated causing the spectrum to tilt In Figure 7 the spectrum is tilted with positive absorbance at lower wavenumbers and negative absorbance at hig...

Страница 10: ...perform an annual examination of your system s health data remotely to determine if any servicing needs to be carried out This includes review of internal diagnostics for warnings and errors component...

Страница 11: ...rease the probability of fracture Prot gez la fibre des contraintes et des courbures serr es rayon de courbure 30 mm interdit Les courbures serr es provoquent une contrainte la surface du verre et aug...

Страница 12: ...en The pen fits over each of the two male fibre ferrules and is aligned by the transparent cover Push and click once on each ferrule Figure 12 Using the fibre cleaning pen to the Souriau plug Once bot...

Страница 13: ...Page 12 DOC1049R 22 February 2024 12 07 www keit co uk Back to Table of Contents Figure 13 Reconnecting the Souriau connectors Locknut Align keyways...

Страница 14: ...ions can cause damage to the instrument If in doubt please consult with Keit Le stockage du spectrom tre IRmadilloNA non aliment et non purg dans des conditions inad quates peut endommager l instrumen...

Страница 15: ...t force we recommend to tighten the thumbscrew until resistance is felt 1 turn For volatile substances it is recommended to keep the stopper cap in place during use Figure 14 Sample cell fitted to IRm...

Страница 16: ...s is an alternative to installing the IRmadilloNA into a main pipeline It is fixed in place using a tri clamp that locates over the rear of the flow cell and two semi circular collars that fit around...

Страница 17: ...ency Annual 4 2 2 Low flow cell ASM1222 The low flow cell that can be secured to the end of the probe for continuous analysis of your process The flow cell is equipped with two 1 16 NPT fittings to at...

Страница 18: ...ll have not perished They should show no signs of swelling tearing or degradation Also inspects the clamp and C pieces for any signs of degradation cracks or damage If the clamp and or the C pieces sh...

Страница 19: ...r as a bench top instrument for offline sampling Angled at 35 the probe can be positioned into beakers or other vessels for ease of use It can also be configured for two heights to give added flexibil...

Страница 20: ...meter over such that the mounting holes are pointed upward the instrument should be disconnected from power coms and purge to avoid damage to the cables or twisting Position the inclined stand over th...

Отзывы: