27
PFP7/REV C/01-08
AC supply not available.
Check mains supply.
Fuse(s) blown.
Check fuses and replace if necessary.
Check the setting of the voltage selector.
7.7
No reading on display
Possible cause
Solution
Incorrect filter selected.
Select correct filter.
Calibration standard incorrectly made.
Re-make standard.
Nebuliser completely blocked.
Check the nebuliser performance as defined in
Section 6.2, Weekly maintenance.
7.8
Unable to set blank
Possible cause
Solution
Blank contaminated.
Re-make the blank and re-check.
Burner encrusted with salt.
Clean the burner as defined in Section 6.4, Six
monthly maintenance.
Ensure that the cleaning procedure is carried out
more frequently.
Ignition electrode encrusted with salt.
Clean or replace the ignition electrode as detailed in
Section 9.1.2.