15
PFP7/REV C/01-08
4.1.2
Rear panel controls
Figure 4.1.2
: Rear panel
Voltage
selection
A two-position switch marked
230 and 115. These positions allow
operation from voltage supplies 190-250 volts and 90-125 volts
respectively at either 50 or 60Hz.
WARNING: when adjusting the operating voltage it may be
necessary to change the top fuse as indicated on the rear panel.
Fuses
Two fuses are fitted to the instrument.
FS1
is the primary fuse provided to
give protection to the instrument and any accessories connected to the
power out
sockets.
FS2
is to protect the electronic circuitry in the PFP7.
Both FS1 and FS2 should be of an anti-surge type.
POWER IN
A three pin receptacle for the AC mains supply
POWER OUT
Two three pin sockets for flame photometer accessories such as a
compressor and a pen recorder. The power from these sockets is
controlled by the instrument
power
switch on the front panel.
RECORDER
Two 4mm sockets which will provide an analogue signal of approximately
1 volt when the readout is 1000 digits or 100.0, 10.00 and 1.000.
FUEL GAS
1/4” connector for fuel tubing.
IMPORTANT: Fuel supply must be regulated to 10-15 inches water
gauge (0.36-0.54psi or 0.025-0.38kg/cm
2
). See section 2.1.2 for
information on different fuel types.
AIR
5/16” connector for air tubing.
IMPORTANT: Air pressure on the inlet to the instrument must not
exceed 30psi (2kg/cm
2
). See section 2.1.3.