Trouble-shooting Registration and Telemetry 

Green light is not illuminating when RAPT Pill plugged in to USB-C 

This occurs if the battery is below a minimum charge level. Please leave the Pill connected for 30 
minutes to gain sufficient charge, then unplug and reconnect. 
 

Red (battery) light flashes rapidly when plugged in to USB-C 

Please install the battery or check that it is making good contact with the terminals. 
 

Lights turn off immediately when disconnected from USB-C 

Please check the charge level of the battery 

 it may be insufficient to power the Pill. The battery 

may be faulty, or not making good connection to the terminals. A fully charged battery has around 
4.2V 

 this can be tested with a multimeter. If the battery has a good charge level, gently bend the 

terminals out in order to ensure there is good contact with the battery. 
 

Battery level reads 0%/not displaying in the RAPT Portal 

The battery needs to reach a full charge when installed in the Pill and connected via USB-C. If the Pill 
has not determined the fully charged state then it will not display the correct percentage. This will 
normally rectify itself after the Pill is used and then fully recharged. 
 

The lights on the Pill stop flashing shortly after it is unplugged 

This is normal behaviour. There is an approximately 10 minute timeout after the Pill is unplugged. 
 

There are no lights on the Pill when unplugged 

This is normal. The Pill will be ‘asleep;

 by default and will wake briefly at the set interval to send 

telemetry. The green light will flash briefly when this happens. 
 

Cannot access the Pill Captive Portal 

This can only be accessed when the Pill is plugged in to USB-C. Access to the portal will cease after 
10 minutes and the Pill will enter a sleep state. Unplugging and reconnecting to USB-C will allow it to 
be reconnected. 
 

User code is not appearing on the Pill Captive Portal 

If the space for the user/validation code on the Pill Captive Portal is blank then make sure that your 
RAPT Pill is connected to your home Wi-Fi network (See 

Determine whether your RAPT Pill is 

connected to Wi-Fi

 below). 

No telemetry is being displayed on the RAPT Portal. 

The default setting for telemetry is 60 minutes. As a result, you will not see telemetry in the RAPT 
Portal right away. Should the telemetry data not appear after 60 has passed you should check the 
following: 
 

Determine whether your RAPT Pill is connected to Wi-Fi 

 
If your RAPT Pill is not connected to your home Wi-Fi network it will be unable to send telemetry to 
the RAPT Portal.  
 
To check this, plug your RAPT Pill in to power and open the RAPT Pill Captive Portal and Navigate to 

Wi-Fi

. Check that it shows that the RAPT Pill is connected to your home Wi-Fi network. 

Содержание RAPT Pill

Страница 1: ...APT Pill Hydrometer Thermometer Quick Start Guide Page 1 of 18 www KegLand com au Last Updated 6 05 2022 16 49 PM RAPT PILL Hydrometer Thermometer KL20596 Quick Start Guide KegLand Distribution PTY LT...

Страница 2: ...vanced Calibration 11 3 Custom Calibration 12 Updating Firmware 14 Trouble shooting Registration and Telemetry 15 Green light is not illuminating when RAPT Pill plugged in to USB C 15 Red battery ligh...

Страница 3: ...KegLand approved cleaners such as StellarOxy and StellarClean and sanitisers such as StellarSan or Phosphoric Acid Battery Installation The battery will come pre installed in the RAPT Pill with a non...

Страница 4: ...ure your battery has good contact with the terminals After installing the battery rotate it slightly within the battery holder to ensure it has good contact Important Once you have installed the Li io...

Страница 5: ...Fully Charged ESP LED GREEN Slow Green Flashes flashes when the captive portal is enabled and the Pill is searching for a Wi Fi network This will happen when the Pill is plugged in to USB C the first...

Страница 6: ...l Captive Portal If you are not prompted to sign into the network of your RAPT Pill by your smart phone or tablet then open your Wi Fi settings and select the RAPT Pill XXXX Wi Fi network again to ope...

Страница 7: ...n be used to access the Pill AP in a browser 2 Ensure you are logged in to the RAPT Portal https app rapt io on a second device that is connected to the same home Wi Fi network as the Pill you are reg...

Страница 8: ...ortant to insert the circuit board in the correct orientation to ensure the correct weight distribution for accurate readings The yellow end of the circuit board should be inserted into the yellow hou...

Страница 9: ...on such as if you are fermenting in a fermentation chamber place the Pill in a jug in the fermentation chamber to confirm that telemetry is received and your RAPT Pill has good signal strength 2 Once...

Страница 10: ...ngs and select the RAPT Pill XXXX Wi Fi network again to open the Pill Captive Portal 3 Once the RAPT Pill Access Portal is open on your phone or tablet navigate to the Calibration tab 4 Select the de...

Страница 11: ...tion is also quite quick and easy and can provide a more accurate result than Standard Calibration It is the same procedure as standard calibration but in addition to water at 1 000 have a solution of...

Страница 12: ...e gravity readings 1 Place the Pill in the first solution of known gravity Wait for the angle to stabilise then make a note of the angle and the gravity 2 Repeat this step for solutions of different g...

Страница 13: ...o this y 0 0000023109x5 0 0005741632x4 0 0568746764x3 2 8119846525x2 71 1087310405x 265 6578215178 This is a sample only DO NOT USE THIS EQUATION AS IT WILL NOT BE ACCURATE Substitute x in this equati...

Страница 14: ...ur phone or tablet or through the Pill IP address 1 Plug your Pill in to USB power with a USB C cable 2 Connect to the Pill Access Portal with your phone or tablet or through the Pill IP address 3 Nav...

Страница 15: ...e Pill stop flashing shortly after it is unplugged This is normal behaviour There is an approximately 10 minute timeout after the Pill is unplugged There are no lights on the Pill when unplugged This...

Страница 16: ...o 30dBm Maximum signal 50dBm Excellent signal 67dBm Good reliable signal 70dBm Ok not a strong signal 80dBM Poor unreliable signal 90dBM Unusable Re Register Your RAPT Pill If your RAPT Pill is still...

Страница 17: ...perating level the Pill will enter a deep sleep mode in order to protect the battery There may have been an interruption to the network or internet The battery may be slightly loose in the terminals c...

Страница 18: ...support please contact your nearest distributor The RAPT Pill Thermometer comes with a 12 month warranty when sold in Australia To lodge a warranty claim in Australia please forward as many visual pie...

Отзывы: