RAPT Pill Hydrometer & Thermometer Quick Start Guide 

 

Page 

12

 of 

18 

www.KegLand.com.au

 

Last Updated 6/05/2022 16:49 PM

 

 

3: Custom Calibration 

Warning! Custom Calibration is for advanced users only

This mode can provide the most accurate readings across a wider range than the other calibration 
modes. 

Only proceed if you are comfortable with spreadsheets and complex curves. If you are working with 
solutions with gravities higher than 1.080, then custom calibration is advised. 

For advanced calibration, you will need to take readings at various gravity points roughly 10 gravity 
points apart, and make a note of the angle of the Pill at each point. We recommend readings from 
1.080 down to 1.000, 10 points apart (so, 1.080, 1.070, 1.060 etc). Note that it does not matter if the 
gravity readings are not exactly 10 points apart, as long as the gravity reading is correct. Use of a 
hydrometer that is known to be accurate is quite important. 

To perform Custom Calibration, it is easiest to start with the strongest solution and dilute to 
progressively weaken the solution. It is important to use a calibrated hydrometer to ensure accurate 
gravity readings. 

1.

 

Place the Pill in the first solution of known gravity. Wait for the angle to stabilise, then make 
a note of the angle and the gravity. 

2.

 

Repeat this step for solutions of different gravity, making a note of the angle and the gravity 
at each step. It is easiest to dilute the starting solution in steps of approximately 10 points. 
Finish with a reading in a solution with a gravity of 1.000 (IE water). 

Once you have sufficient readings, the angle and the gravity need to be entered on an Excel 
spreadsheet (or similar such as Google Sheets) as a graph wit

h gravity on the ‘y’ axis and angle on 

the ‘x’ axis. The more readings you take, the more accurate this calibration will be. We recommend a 

minimum of 6 samples, with 10 or more providing a greater degree of accuracy.  

When you have created the graph, create a polynomial trend line onto the graph with the degree of 
the polynomial being 

at least

 one less than the number of data points. If the trend line creates an 

irregular curve (such as not passing through the data point at 1.000 correctly or not following a 
polynomial shape) reduce the number of data points (ensuring that there are still more data points 
than the degree of the polynomial) until the curve matches. 

With 10 data points, a polynomial degree between 5 and 8 will typically give the best results. 

Содержание RAPT Pill

Страница 1: ...APT Pill Hydrometer Thermometer Quick Start Guide Page 1 of 18 www KegLand com au Last Updated 6 05 2022 16 49 PM RAPT PILL Hydrometer Thermometer KL20596 Quick Start Guide KegLand Distribution PTY LT...

Страница 2: ...vanced Calibration 11 3 Custom Calibration 12 Updating Firmware 14 Trouble shooting Registration and Telemetry 15 Green light is not illuminating when RAPT Pill plugged in to USB C 15 Red battery ligh...

Страница 3: ...KegLand approved cleaners such as StellarOxy and StellarClean and sanitisers such as StellarSan or Phosphoric Acid Battery Installation The battery will come pre installed in the RAPT Pill with a non...

Страница 4: ...ure your battery has good contact with the terminals After installing the battery rotate it slightly within the battery holder to ensure it has good contact Important Once you have installed the Li io...

Страница 5: ...Fully Charged ESP LED GREEN Slow Green Flashes flashes when the captive portal is enabled and the Pill is searching for a Wi Fi network This will happen when the Pill is plugged in to USB C the first...

Страница 6: ...l Captive Portal If you are not prompted to sign into the network of your RAPT Pill by your smart phone or tablet then open your Wi Fi settings and select the RAPT Pill XXXX Wi Fi network again to ope...

Страница 7: ...n be used to access the Pill AP in a browser 2 Ensure you are logged in to the RAPT Portal https app rapt io on a second device that is connected to the same home Wi Fi network as the Pill you are reg...

Страница 8: ...ortant to insert the circuit board in the correct orientation to ensure the correct weight distribution for accurate readings The yellow end of the circuit board should be inserted into the yellow hou...

Страница 9: ...on such as if you are fermenting in a fermentation chamber place the Pill in a jug in the fermentation chamber to confirm that telemetry is received and your RAPT Pill has good signal strength 2 Once...

Страница 10: ...ngs and select the RAPT Pill XXXX Wi Fi network again to open the Pill Captive Portal 3 Once the RAPT Pill Access Portal is open on your phone or tablet navigate to the Calibration tab 4 Select the de...

Страница 11: ...tion is also quite quick and easy and can provide a more accurate result than Standard Calibration It is the same procedure as standard calibration but in addition to water at 1 000 have a solution of...

Страница 12: ...e gravity readings 1 Place the Pill in the first solution of known gravity Wait for the angle to stabilise then make a note of the angle and the gravity 2 Repeat this step for solutions of different g...

Страница 13: ...o this y 0 0000023109x5 0 0005741632x4 0 0568746764x3 2 8119846525x2 71 1087310405x 265 6578215178 This is a sample only DO NOT USE THIS EQUATION AS IT WILL NOT BE ACCURATE Substitute x in this equati...

Страница 14: ...ur phone or tablet or through the Pill IP address 1 Plug your Pill in to USB power with a USB C cable 2 Connect to the Pill Access Portal with your phone or tablet or through the Pill IP address 3 Nav...

Страница 15: ...e Pill stop flashing shortly after it is unplugged This is normal behaviour There is an approximately 10 minute timeout after the Pill is unplugged There are no lights on the Pill when unplugged This...

Страница 16: ...o 30dBm Maximum signal 50dBm Excellent signal 67dBm Good reliable signal 70dBm Ok not a strong signal 80dBM Poor unreliable signal 90dBM Unusable Re Register Your RAPT Pill If your RAPT Pill is still...

Страница 17: ...perating level the Pill will enter a deep sleep mode in order to protect the battery There may have been an interruption to the network or internet The battery may be slightly loose in the terminals c...

Страница 18: ...support please contact your nearest distributor The RAPT Pill Thermometer comes with a 12 month warranty when sold in Australia To lodge a warranty claim in Australia please forward as many visual pie...

Отзывы: