APPLANATION TONOmETER ‘KEELER FIxED’ (R-TyPE)
This instrument is for those who wish the tonometer to
remain permanently on the slit lamp.
• Mount the plate for the tonometer onto the
microscope body using the securing screw.
• Mount the tonometer mount onto the
mounting post.
• Swing the Tonometer forward in front of the
microscope for examination. A notch position
ensures exact centring of the prism with the
left objective.
• To obtain an image as clear and as
free of reflexes as possible, the angle
between the illumination and the
microscope should be about 60° and
the slit diaphragm should be fully
opened.
• When not in use the instrument is
swung around and secured in a notch
position to the right of the microscope.
10.4 fitting applanation tonometerS, t type and r typeS
KEELER APPLANATION TONOmETER (T-TyPE)
• Position the guide plate in the tonometer/test bar
support hole on the slit lamp.
• Lift Tonometer out of the packaging and assemble
it by inserting the pin on its base into one of the
two possible openings (for right or left eye) on the
horizontal guide plate above the slit lamp axis.
These positions are related to the microscope
optics and observation can be made either
through the right or the left eye-piece.
• The tonometer will slip easily onto the support
plate; stability is assured by the locking pins.
• To obtain an image as clear and as
free of reflexes as possible, the angle
between the illumination and the
microscope should be about 60° and the
slit diaphragm should be fully opened.
• When not in use the Tonometer should
be removed from the Slit Lamp and
placed securely back in the packaging or
a suitable location.
Right eye
Left eye