14

S L I T   L A M P   b y   K E E L E R

www.keeler.co.uk

11.2 preparing the patient and uSing the Slit lamp

1

   The patient should be as comfortable as possible and with the patient 

in the chinrest adjust the chinrest height (7) so that the patient’s eyes 
are level with the height marking (4) on the chinrest support.

2  

Focus the eyepieces using the test bar as described earlier, and if you 
have not already done so set them to your interpupillary distance by 
holding both eyepiece bodies and rotating  them inwards or outwards 
until they are correct for your PD.

3  

Switch on the illumination, making sure the rheostat (25) is set to a low 
level to minimise the patient’s exposure to light hazard. 

Rotate the joystick (24) until the light beam is at eye level.

5  

Holding the joystick vertical, move the slit Lamp base towards the 
patient until the slit beam appears focused on the patient’s cornea.

6  

Adjust slit width (18), magnification (31), slit rotation (13) & slit angle 
etc. as required to perform the examination. 

7  

Loosening slit offset centring knob (16) to allow the slit image to be 
moved off centre for scleral illumination. Tightening the knob will re-
centre the slit image in the centre of the visual field of the microscope. 

8  

The slit image is made vertical, or given a preset angle by means of the 
illumination latch (17) (notches at 5°, 10° and 15° & 20°).

9  

When using the blue filter (12) the user may wish to insert the yellow 
barrier filter (29). The yellow barrier filter is out when the knob is up, in 
when it is down.

10  

When the examination is complete, set the rheostat to a low level and 
switch off the Slit Lamp.

Shut down after every use. In case the dust cover is 
used: risk of overheating.

Parts of the slit lamp coming in to contact with the patient 
should be cleaned in accordance with these instructions prior  
to the examination. Keeler recommends the use of disposable 
hygienic chinrest tissues on the Chinrest before the patients 
place their chin on it.

Never use the instrument if visibly damaged and periodically 
inspect it for signs of damage or misuse.

7

4

25

24

29

31

12

13

16
17

18

Содержание Slit Lamp H Series

Страница 1: ...Slit Lamp Instructions for use H Series ...

Страница 2: ...n whole or part without the manufacturer s prior written approval As part of our policy for continued product development we the manufacturer reserve the right to make changes to specifications and other information contained in this document without prior notice Copyright Keeler Limited 2013 Published in the UK 2013 ...

Страница 3: ... Transport storage and working conditions Names of controls and components 8 Please click on the contents to go straight to your chosen section or navigate by using the Next and Back buttons to the right Clicking on Home will bring you back to this page Assembly 9 Instructions for use 13 Description of filters apertures and magnifications 15 Routine maintenance 16 Warranty 17 Specifications and el...

Страница 4: ... uk 1 Introduction Thank you for choosing this Keeler product Please read this manual carefully before using your Keeler Slit Lamp this will ensure the safety of the patient and ensure you get the best performance from this precision optical device ...

Страница 5: ...e B Applied Part Non ionizing radiation Keep dry Reference Number Serial Number Date of manufacture The CE mark on this product indicates it has been tested to and conforms with the provisions noted within the 93 42 EEC Medical Device Directive This way up Material suitable for recycling Fragile The symbol on the Product or on it s Packaging and instructions indicates it was put on the market plac...

Страница 6: ...teristics of the examination light The Keeler Slit Lamp is designed and built in conformity with EC Directive 93 42 EEC and the ISO 9000 and ISO 13485 series of quality standards The CE European Community mark attests that the Keeler Slit Lamp complies with the provisions of the EC Directive 93 42 EEC Classification CE Regulation 93 42 EEC Class I FDA Class II IEC EN Standard 60601 1 H series Safe...

Страница 7: ...d eyes will be at greater risk The risk may also be increased if the person being examined has had any exposure with the same instrument or any other ophthalmic instrument using a visible light source during the previous 24 hours This will apply particularly if the eye has been exposed to retinal photography It is well established that exposure of the eye to intense light sources for extended peri...

Страница 8: ...liminate risk of tripping or damage to user Before any cleaning of the instrument or the base unit ensure the power lead is disconnected Bulbs can reach high temperatures in use allow to cool before handling Do not exceed maximum recommended exposure time Should the instrument suffer shocks for example should it accidentally fall and the optical system or the illumination system are damaged it may...

Страница 9: ...he instrument Use caution to ensure cloth is not saturated with solution c Surfaces must be carefully hand dried using a clean non shedding cloth d Safely dispose of used cleaning materials 8 Transport storage and working conditions The following ambient condition limits are recommended for the Keeler Slit Lamp for transport and storage it is recommended that the Slit Lamp is kept in its original ...

Страница 10: ...tion mirror 16 Slit offset centring knob 17 Inclination latch 5 to 20 18 Slit width controls 19 Test bar tonometer plate mounting hole and cover 20 Illumination arm locking knob 21 Microscope arm locking knob 22 Eyepiece assembly securing knob 23 Joystick base locking knob 24 Joystick control X Y Z movement 25 Illumination control rheostat 26 Axle 27 Runner covers 28 R type tonometer mounting hole...

Страница 11: ...tors ensure the following before moving it to another location a The table is at it s lowest position b The power cord is removed c The slit lamp arm and base locking knobs are tightened d The Runner covers are securely located e The system is moved by grasping it at its lowest practical point 10 1 Table Top and Base Assembly procedure 1 Attach the Slit Lamp table top to your table leg using the M...

Страница 12: ... Connect the main lamp cable from the Chinrest to the illumination tower Do not twist the lead on the tower illumination system 2 Connect the power cables a Chinrest fixation light cable to power supply unit b 3 pin cable from power supply unit to slit lamp base assembly c 4 pin main lamp cable from bottom of the chinrest to the slit lamp base assembly d Ensure cables are routed to allow free move...

Страница 13: ...d provided Only a hospital grade 3 conductor electrical power supply cable must be used For USA and Canada Detachable power supply cord set UL listed type SJE SJT or SJO 3 conductor not smaller than 18 AWG Plug cable and ground lead connection of the socket have to be in perfect condition 3 pin to 3 pin Fixation light socket ...

Страница 14: ... the microscope 10 4 Fitting Applanation Tonometers T type and R types Keeler Applanation Tonometer T Type Position the guide plate in the tonometer test bar support hole on the slit lamp Lift Tonometer out of the packaging and assemble it by inserting the pin on its base into one of the two possible openings for right or left eye on the horizontal guide plate above the slit lamp axis These positi...

Страница 15: ...d rotating them inwards or outwards until they are correct for your PD 4 Turn both eyepieces 32 to maximum plus correction 5 Close one eye and with the other eye look through the microscope slowly turning the open eye eyepiece towards the minus position until the image of the test bar is in focus Stop 6 Repeat the above process for the other eyepiece 7 Make a note of the positions of the eyepieces...

Страница 16: ...on 13 slit angle etc as required to perform the examination 7 Loosening slit offset centring knob 16 to allow the slit image to be moved off centre for scleral illumination Tightening the knob will re centre the slit image in the centre of the visual field of the microscope 8 The slit image is made vertical or given a preset angle by means of the illumination latch 17 notches at 5 10 and 15 20 9 W...

Страница 17: ...view 6x 34mm 10x 22mm 16x 14mm 25x 8 5mm 40x 5 5mm Filters An Infra Red IR cut filter is permanently fitted for safety reasons a Clear b Neutral Density c Red Free d Diffuser e Blue Apertures Aperture diameters mm Tower illumination Tower has the facility to tilt towards the user and positively locates at each step 0 5 10 15 and 20 a b c d e 0 2 1 2 3 5 9 12 12 Description of filters apertures and...

Страница 18: ...ignificant life expectancy we suggest that the Slit Lamp is always switched off between examinations to conserve energy and LED life In the unlikely event of an LED failure please contact Keeler or your local distributor for guidance on the replacement procedure 13 2 Regularly inspect the device for damage or dirt Routinely clean as per section 7 Cleaning Instructions 13 3 Cleaning and changing th...

Страница 19: ... Slit Lamp becomes hard to move on the glide plate the plate should be cleaned with a lightly oiled cloth or silicon polish The axle should be cleaned only with dry lint free cloths The Keeler H Series Slit Lamps are guaranteed for three years 3 against faulty workmanship materials or factory assembly Warranty is on a Return To Base RTB basis at the cost of the customer and may be void if the Slit...

Страница 20: ...below The customer or user should ensure that it is used in such an environment Immunity test IEC 60601 Test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 6100 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air The Keeler Slit Lamp uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interfe...

Страница 21: ...ed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the Keeler Slit Lamp is used exceeds the applicable RF compliance level above the Keeler Slit Lamp should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orientating or relocating the Keeler Slit La...

Страница 22: ...lit Lamp 40H Standard Set complete with Table Top Slit Lamp PSU 3020 P 2000 3020 P 5032 3020 P 5040 Slit Lamp 40H Refraction Unit Set Slit Lamp Power Supply and Socket Kit 3020 P 2003 3020 P 5032 3020 P 7017 Slit Lamp 40H LED Standard Set complete with Table Top LED Slit Lamp PSU 3020 P 2007 3020 P 5056 3020 P 5040 Slit Lamp 40H LED Refraction Unit set LED Slit Lamp Power Supply and Socket Kit 302...

Страница 23: ...ection mirror EP39 80250 Long projection mirror standard EP39 80052 Dust cover EP39 80273 Part name Part number Chinrest papers 3104 L 8200 Chinrest assembly 3020 P 5036 Power supply to Slit Lamp base cable Table top variant only 3020 P 7011 Base assembly Table top variant only 3020 P 5007 Refraction Stand cable Kit R Stand variant only 3020 P 7014 Joystick rubber EP39 70369 ...

Страница 24: ...mber Mains Cables EU MIS138 Mains Cables Brazil 3020 P 7007 Mains Cables Japan 3020 P 7008 Mains Cables UK MIS100 Mains Cables USA 3020 P 7016 Mains Cables Australia 3020 P 7022 Mains Cables China 3020 P 7023 Part name Part number Volk digital lens set 2105 L 2010 Tonometer R type 2414 P 2040 Tonometer T type 2414 P 2030 ...

Страница 25: ...0 51261868 Fax 86 10 58790155 Disposal of old Electrical and Electronic Equipment Applicable in the European Union and other European Countries with separate Collection Systems This Symbol on the Product or on its Packaging and instructions indicates that it was put on the market place after August 2005 and that this product shall not be treated as Household Waste To Reduce the Environmental impac...