9
S L I T L A M P b y K E E L E R
www.keeler.co.uk
Your Keeler Slit Lamp has been designed to fit on to an electric insulated
medical table base or on to an electric insulated and fire resistant medical
table top, e.g. a Refraction Stand or Combi Unit.
Take care when unpacking your Slit Lamp that you do not accidentally
damage or discard any of the contents.
Leave the Slit Lamp in the packing for several hours after
delivery before unpacking to reduce the risk of condensation
forming.
Keeler Slit Lamps can be fitted to most Refraction Stands / Combi
units. Keeler recommends that this be carried out by suitably trained
technicians to ensure performance and safety are not compromised.
The Refraction Stand, Combi Unit or table leg must be 60601-1
3rd edition compliant
If you are fitting or have fitted your Slit Lamp to a medical or Keeler table
leg/base ensure it is situated on a firm and level floor.
If the table leg/base has castors ensure the following before moving it to
another location:
a) The table is at it’s lowest position
b) The power cord is removed
c) The slit lamp arm and base locking knobs are tightened
d) The Runner covers are securely located
e) The system is moved by grasping it at its lowest practical point.
10.1 table top and baSe aSSembly procedure
1
Attach the Slit Lamp table top to your table leg using the M6 x 20mm
CAP head fasteners and washers. Note that the power supply and
accessory drawer should face the operator.
The security of the fitment of the table top to the table leg is
critical for patient and Slit Lamp safety
2
Using the wrench provided, fit the Headrest Assembly to the tabletop
using the hex bolts and washers. The Headrest Assembly locates on the
underside of the tabletop. Take care not to over tighten the hex bolts.
3
Attach the Patient Grab Handles (8) to the Headrest Assembly.
4
Place the Slit Lamp base on the Runners. Ensure that the wheels are in
line with each other. Check that the guide wheels are tight.
5
Fit the Runner covers to the Runners by gently sliding them inwards,
towards each other.
10 aSSembly
8
Runners