12. After each reading is taken, the display shows the average of all the
readings taken on the eye. When four or more readings have been
taken the average displayed is the IOP recorded for the patient.
13. You can review the readings taken using the
Review
button. If a non
event or bad event has been detected
ER
appears in the display when
reviewing.
Note:
Each measurement is taken to 1 decimal place but is displayed to the
nearest whole number; 0.5 would be rounded up to 1 mm Hg.
The Next Eye
Once a satisfactory number of readings have been taken on the right eye
press the
Change Eye
button, the left arrow above the Change Eye
button illuminates. Readings can now be taken on the left eye.
See the
Change Eye Indicators and the Change Eye Button on page 13 for
more details
.
Printing the Results
The Pulsair EasyEye has a built in printer which prints a patient’s record
once all the readings have been taken.
When the hand unit is returned to the holster the results are automatically
printed if the printer is turned on.
If the printer is turned off when the hand unit is returned to the holster,
press the
Printer
on/off switch to turn the printer on, then press the
Change Eye button on the hand unit (do NOT remove the hand unit from
the holster). The last recorded readings are then printed.
The Pulsair EasyEye memory is cleared when the hand unit is removed
from the holster for use on another patient, therefore if a printout is
required you must turn the printer on before removing the hand unit from
the holster for use on another patient.
To print off another copy of the report press the
Change Eye
button
whilst the hand unit is in the holster.
If you want to review the patients’ readings you must do this before the
hand unit is removed from the holster for use on another patient.