Puls
air EasyEye
Eyepiece
The Eyepiece allows the user to view the patients eye and align the
targetting system.
QuickPulse Indicator
This indicator illuminates if the
QuickPulse
mode is active.
QuickPulse Button
QuickPulse mode may need to be invoked for patients who have corneal
scarring, a high level of astigmatism, or a less reflective eye than normal.
The patient should be asked to blink several times and further readings
attempted before engaging QuickPulse.
The Pulsair EasyEye checks that the eye focus is correct and if so,
produces an air pulse.
Before invoking QuickPulse mode ensure that the correct image is seen
through the eyepiece.
If this is the case and no air pulse occurs, it may be that the particular eye
is less reflective than normal. Press the
QuickPulse
button. It should now
be easier to obtain a reading. QuickPulse should always be used on post
operative eyes with corneal scarring. QuickPulse mode can be toggled on
and off by pressing the button, the QuickPulse Indicator illuminates if the
mode is active.
The QuickPulse facility should only be used when necessary.
QuickPulse can also aid in training new operators. They may have difficulty
initially in obtaining the correct focus (refer to the pull out instructions,
located underneath the Pulsair EasyEye).