1 9
EP59-50399
Issue D
English
When the hand unit is removed from the holster the two green LED’s on
the front illuminate, and the pump starts (an audible tone is emitted). The
pump takes 2 seconds to prime before the unit is operational.
Before using the Pulsair EasyEye press the
Demo
button to dispel any
minute particles of dust or moisture which may have settled whilst the
Pulsair EasyEye was not in use.
Taking Readings
For quick reference, please refer to the pull out short form instructions
located underneath the Pulsair EasyEye unit.
Once the Pulsair EasyEye and the patient are prepared, you are ready to
take a reading.
1. The Pulsair EasyEye is set to automatically select the right eye. If you
wish to select the left eye, press the
Change Eye
button on the hand
unit.
2. Lift the hand unit, the pump starts and the two green LED’s illuminate.