Next
Home
Back
9
4. Cleaning and disinfection instruction
4.2 Disinfecting the Tonometer prisms
Always disinfect the Tonometer prisms before use. Hand hygiene
must be considered to prevent any contamination.
1
Carefully remove the
Tonometer prism from
the prism holder.
2
Wash the Tonometer prism
under cold running water for
approximately 1 minute, to
ensure Tonometer prism
is physically clean before
exposed to disinfection
process.
3
Immerse the Tonometer prism
in the disinfectant fluid. Types
of disinfectant fluid vary.
4
Rinse the disinfectant from
the prism in running water
for between 10 and 30
minutes.
5
Dry the disinfected
Tonometer prism with a
clean soft cloth.
6
Store the Tonometer prism in a
suitable container ready for use.
Safely dispose of the
disinfectant fluids used.
Please follow disinfectant solution guidelines for instructions, concentration and
time of immersion. (For example: Pantasept - 3% aqueous solution for 10 minutes,
Hydrogen Peroxide 3% aqueous solution for 10 minutes, Sodium Hypochlorite,
10% aqueous solution for 10 minutes etc.).
Do not disinfect using
the following:
Alcohol, Acetone, UV radiation,
Sterilisation, Immersion in fluid for
more than one hour,
Temperatures greater than 60°C
1 min
10 mins
10-30
mins