5.3  CABLE ATTACHMENT PROCEDURE

1.  Connect the main lamp cable from the chinrest to the illumination tower. 

Do not twist the lead on the tower illumination system.

2.  Connect the power cables.

a)  Chinrest fi xation light cable to power supply unit.
b)  (3 pin) cable from power supply unit to slit lamp base assembly.
c)  (4 pin) main lamp cable from bottom of the chinrest to the slit lamp 

base assembly.

d)  Ensure cables are routed to allow free 

movement of the XYZ base and to be 

clear of the patients.

If your Slit Lamp was not supplied with a transformer (Part #3020-P-5040), 

make sure that the power connection is compatible with the specifi cations 

in this manual and is connected by a qualifi ed technician to an available 

and suitable power supply, see section 9.4 Power Supply page 

20

.

3.  Connect the mains power to the Slit Lamp transformer using the power lead provided.

Only a hospital grade 3-conductor electrical power supply cable must be 

used. For USA and Canada: Detachable power supply cord set, UL listed, 

type SJE, SJT or SJO, 3-conductor, not smaller than 18 AWG. Plug, cable 

and ground lead connection of the socket have to be in perfect condition.

5.4  FITTING APPLANATION TONOMETERS, T TYPE AND R TYPES

Keeler Applanation Tonometer (T-TYPE)

1.  Position the guide plate in the tonometer / test bar support hole on the slit lamp.
2.  Lift the tonometer out of the packaging and assemble it by inserting the pin on its base 

into one of the two possible openings (for right or left eye) on the horizontal guide 

plate above the slit lamp axis. These positions are related to the microscope optics and 

observation can be made either through the right or the left eye-piece.

Cable connector

4 pin
3 pin

3 pin to 3 pin

Fixation light socket

Содержание H Series

Страница 1: ...SLIT LAMP INSTRUCTIONS FOR USE H Series A world without vision loss...

Страница 2: ......

Страница 3: ...R USE 11 6 1 SETTING THE BINOCULARS 11 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP 12 6 3 DESCRIPTION OF FILTERS APERTURES AND MAGNIFICATIONS 13 7 ROUTINE MAINTENANCE 14 7 1 LED SYSTEMS 14 7 2 R...

Страница 4: ...ty Humidity limitation Use by date Serial number Catalogue number Medical device Translation The Keeler Slit Lamp is designed and built in conformity with Directive 93 42 EEC Regulation EU 2017 745 an...

Страница 5: ...embly and a table top with power supply and accessory drawer The light intensity is controlled by a variable rheostat located on the XYZ translation base There are a number of selectable filters allow...

Страница 6: ...the specialized technicians of the manufacturer s Technical Service Centre or by personnel trained and authorised by the manufacturer The manufacturer declines any and all responsibility for loss and...

Страница 7: ...ortant when the unit has been stored or transported in a cold environment this can cause severe condensation to develop on the optical elements 2 3 CONTRAINDICATION There is no restriction to patient...

Страница 8: ...s 9 Power lead chinrest to lamp housing KSL H Series 10 Lamp cover 11 Lamp cover release screws 12 Lever for grey ND blue diffuser and red free filters 13 Slit length slit rotation and aperture contro...

Страница 9: ...8 R type tonometer mounting hole 29 Yellow filter knob up out 30 Lock for securing magnification body 31 Magnification change drum 32 Eyepieces adjustable for PD and dioptre correction 33 Breath shiel...

Страница 10: ...castors ensure the following before moving it to another location 1 The table is at its lowest position 2 The power cord is removed 3 The Slit Lamp arm and base locking knobs are tightened 4 The runn...

Страница 11: ...gned Attach the tower to the base using the hex bolt removed earlier and tighten using the wrench provided 2 Carefully fit the microscope body to the arm ensuring it is pushed to the stop Tighten usin...

Страница 12: ...n 9 4 Power Supply page 20 3 Connect the mains power to the Slit Lamp transformer using the power lead provided Only a hospital grade 3 conductor electrical power supply cable must be used For USA and...

Страница 13: ...sures exact centering of the prism with the left objective 8 To obtain an image as clear and as free of reflexes as possible the angle between the illumination and the microscope should be about 60 an...

Страница 14: ...nus one 1 or minus two 2 dioptres 6 2 PREPARING THE PATIENT AND USING THE SLIT LAMP Parts of the slit lamp coming in to contact with the patient should be cleaned in accordance with these instructions...

Страница 15: ...mination Tightening the knob will re centre the slit image in the centre of the visual field of the microscope 8 The slit image is made vertical or given a pre set angle by means of the inclination la...

Страница 16: ...e outlined below should only be carried out once the main power cable has been disconnected If you have any problems that are not covered by the procedures described below contact Keeler Ltd or your l...

Страница 17: ...ables and connectors To access the bulb connections see earlier in this section for guidance 7 5 OPTICS The optics should be wiped clean of any loose dirt or debris with a suitable dust brush then cle...

Страница 18: ...g or using this instrument please read carefully and observe what is described here Portable or mobile type radio frequency communication units may have an adverse effect on this instrument resulting...

Страница 19: ...1 kV line s to line s 2 kV line s for input output line s 1 kV line s to line s 2 kV line s for input output line s Mains power quality should be that of a typical professional healthcare facility Vol...

Страница 20: ...nd people 1 Field strengths from fixed transmitters such as base stations cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predict...

Страница 21: ...requency applies These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 9 4 TECHNICAL SPECIFICATIONS O...

Страница 22: ...provide at least two means of protection Power Supply Power supply unit Switch mode 100V 240V input 10 multi plug compliant to EN IEC 60601 1 EN IEC 61000 6 2 EN IEC 61000 6 3 Fuse T2 5AH 250V Power s...

Страница 23: ...60 hPa 700 hPa TRANSPORT CONDITIONS 50 C 40 C 95 10 1060 hPa 500 hPa Vibration sinusoidal 10 Hz to 500 Hz 0 5g Shock 30 g duration 6 ms Bump 10 g duration 6 ms This instrument does not meet the temper...

Страница 24: ...al Keeler Applanation Tonometer 2414 P 2032 Keeler D KAT R Type Digital Keeler Applanation Tonometer 2414 P 2042 11 PACKAGING AND DISPOSAL INFORMATION Disposal of old electrical and electronic equipme...

Страница 25: ......

Страница 26: ...2021 USA Sales Office Keeler USA 3222 Phoenixville Pike Building 50 Malvern PA 19355 USA Toll Free 1 800 523 5620 Tel 1 610 353 4350 Fax 1 610 353 7814 India Office Keeler India Halma India Pvt Ltd P...

Отзывы: