5.3 CABLE ATTACHMENT PROCEDURE
1. Connect the main lamp cable from the chinrest to the illumination tower.
Do not twist the lead on the tower illumination system.
2. Connect the power cables.
a) Chinrest fi xation light cable to power supply unit.
b) (3 pin) cable from power supply unit to slit lamp base assembly.
c) (4 pin) main lamp cable from bottom of the chinrest to the slit lamp
base assembly.
d) Ensure cables are routed to allow free
movement of the XYZ base and to be
clear of the patients.
If your Slit Lamp was not supplied with a transformer (Part #3020-P-5040),
make sure that the power connection is compatible with the specifi cations
in this manual and is connected by a qualifi ed technician to an available
and suitable power supply, see section 9.4 Power Supply page
20
.
3. Connect the mains power to the Slit Lamp transformer using the power lead provided.
Only a hospital grade 3-conductor electrical power supply cable must be
used. For USA and Canada: Detachable power supply cord set, UL listed,
type SJE, SJT or SJO, 3-conductor, not smaller than 18 AWG. Plug, cable
and ground lead connection of the socket have to be in perfect condition.
5.4 FITTING APPLANATION TONOMETERS, T TYPE AND R TYPES
Keeler Applanation Tonometer (T-TYPE)
1. Position the guide plate in the tonometer / test bar support hole on the slit lamp.
2. Lift the tonometer out of the packaging and assemble it by inserting the pin on its base
into one of the two possible openings (for right or left eye) on the horizontal guide
plate above the slit lamp axis. These positions are related to the microscope optics and
observation can be made either through the right or the left eye-piece.
Cable connector
4 pin
3 pin
3 pin to 3 pin
Fixation light socket