8
EN
KSL-H by KEELER
5. ASSEMBLY
Your Keeler Slit Lamp has been designed to fit on to an electric insulated medical table base or on
to an electric insulated and fire resistant medical table top, e.g. a refraction stand or combi unit.
Take care when unpacking your Slit Lamp that you do not accidentally damage or discard any
of the contents.
Leave the Slit Lamp in the packing for several hours after delivery before
unpacking to reduce the risk of condensation forming.
Keeler Slit Lamps can be fitted to most refraction stands / combi units. Keeler advises that this
shall be carried out by suitably trained technicians to ensure performance and safety are not
affected.
The refraction stand, combi unit or table leg must be IEC 60601-1
compliant.
If you are fitting or have fitted your Slit Lamp to a medical or Keeler table leg / base ensure it is
situated on a firm and level floor.
If the table leg / base has castors ensure the following before moving it to another location:
1. The table is at its lowest position.
2. The power cord is removed.
3. The Slit Lamp arm and base locking knobs are tightened.
4. The runner covers are securely located.
5. The system is moved by grasping it at its lowest practical point.
5.1 TABLE TOP AND BASE ASSEMBLY PROCEDURE
1. Attach the Slit Lamp table top to your table leg using the M6 x 20mm screws and
washers. Note that the power supply and accessory drawer should face the operator.
The security of the fitment of the table top to the table leg is critical for
patient and Slit Lamp safety
2. Using the wrench provided, fit the headrest assembly to the tabletop using the hex bolts
and washers. The headrest assembly locates on the underside of the tabletop. Take care
not to over tighten the hex bolts.
3. Attach the patient grab handles (8) to the headrest assembly.
4. Place the Slit Lamp base on the runners. Ensure that the wheels are in line with each
other. Check that the guide wheels are tight.
5. Fit the runner covers to the runners by gently sliding them inwards, towards each other.
Runners