3. Safety
3.2 Warnings and Cautions
Observe the following precautions in order to ensure safe operation of
the instrument.
• This device is intended to be used only by suitably trained and
authorised healthcare professionals.
• US Federal Law restricts this device to sale by or on the order
of a physician or practitioner.
• Accuracy of applanation IOP measurements is known to be affected
by variations and changes in corneal rigidity due to differences in
corneal thickness, intrinsic structural factors or corneal refractive
surgery. It is recommended that these factors are considered during
IOP measurement.
• Never use the instrument if visibly damaged. Periodically
inspect it for signs of damage or misuse.
• The contact surface of the prism should be checked before each use
for damage and discarded if damage is found.
• Check your Keeler product for signs of transport / storage damage
prior to use.
• We recommend that the prism is not used when it becomes more
than two years of age as, after this time it is possible that body or
sterilising fluids may seep inside leading to possible sterility and cross
contamination issues.
• Only decontaminate / clean in accordance with a method given in
Cleaning (Section 4) of this instruction manual.
• Use only cloths dampened with water to clean the Applanation
Tonometer body. Do not use corrosive products or alcohol.
• This product should not be immersed in fluid.
• Keeler Applanation Tonometers should be serviced and calibrated
annually. Any servicing or repairs / modifications should only be
carried out by Keeler Ltd. or by suitably trained and authorised
distributors. The manufacturer declines any and all responsibility for
loss and / or damages resulting from unauthorised repairs;
furthermore, any such actions will invalidate the warranty.
• Never use the instrument if the ambient temperature, atmospheric
pressure, and/or relative humidity are outside the limits specified in
this manual.
• Should the instrument suffer shocks (for example, should it
accidentally fall), follow the check procedure outlined in the
'Servicing and Calibrations' (Section 12); if necessary, return the
instrument to the manufacturer for repair.
• Use only the listed accessories in conjunction with the instrument;
use said accessories only in accordance with the procedures set forth
in the instruction manuals.
• Always carefully observe the safety rules and other precautions
published herein.
• Do not use in the presence of flammable gases / liquids, or in an
oxygen rich environment.
• For indoor use only (protect from moisture).
• Electrical equipment can be affected by electromagnetic interference.
If this occurs whilst using this equipment, switch the unit off and
reposition.
• If during measurement the slit lamp is moved forward toward the
patient or the patient moves toward the slit lamp, the sensor arm will be
pushed into contact with a stop spring.