9. Problem solving
9.1 The fluorescein ring is too wide or too narrow
Home
Back Next
Cause:
The fluorescein semicircles are too wide.
The measuring prism was not dried after
cleaning, or the eyelids came into contact
with the measuring prism during
measurement. The pressure reading is higher
than the real intraocular pressure.
Correction:
Move the slit lamp back and dry the
measuring prism with a wad of sterile
cotton wool or lint free cloth.
Cause:
The fluorescein semicircles are too narrow.
The lacrimal fluid has dried during prolonged
measurement. The pressure reading is lower
than real ocular pressure.
Correction:
Move the Slit Lamp back and ask the
patient to close his / her eyes once or twice,
then repeat the measurement procedure.
Continued