7. Switching on / off and configuring the D-KAT
7.1 Switching on / off
To switch the D-KAT on, press the function button briefly and
the display will show a nominal reading. The D-KAT can be
switched off by pressing the function button again. The unit
will power down when not in use after 3 minutes.
7.2 Changing decimal point setting
Turn on the D-KAT and press the function button for more
than 3 seconds. The display will flash briefly. Release the
function button “dP” (decimal point) or “ndP” (no decimal
point) will be displayed. Press the function button briefly to
toggle between both. Once the desired display has been set
release the function button and the D-KAT will return to the
pressure readings display.
7.3 Changing the display brightness
Turn on the D-KAT and press the function button for more
than 3 seconds. The display will flash briefly. Release the
function button “dP” (decimal point) or “ndP” (no decimal
point) will be displayed.
Press the function again for more than 3 seconds, the
display will flash and once the function button is released,
the display will show "L" together with 0,1,2 or 3. Press
the function button briefly to toggle to the desired display
brightness.
Note: The D-KAT will return to the main display at any time if the
function button is not pressed within the timings listed above.
7.4 Battery status
If “bat” appears in the display when switching on instead
of a numerical value, the battery should be replaced
immediately. The battery should be replaced using a 1.5v AA
Alkaline type.