4. Safety
Caution
• Use only genuine Keeler approved parts and accessories or
device safety and performance may be compromised
• The product has been designed to function safely when at an
ambient temperature b10°C and +35°C
• Keep out of the reach of children
• To prevent condensation from forming, allow instrument to
come to room temperature before use
• For indoor use only (protect from moisture)
• Keep the console away from sources of liquids and do not
spray with water
• This product is suitable for use with only Nitrous Oxide or
Carbon Dioxide medical gases
• Only non-syphonic gas cylinders should be used with this device
• No modification of this equipment is allowed
• Care should be taken not to trap fingers in pinch points
during gas bottle change
• Follow guidance on cleaning / routine maintenance to
prevent personal injury / damage to equipment
Switch off the electrical supply and disconnect from the
mains electrical supply before cleaning and inspection
• Do not use hypercarbonate or phenolic based cleaning
solutions or disinfectants containing cationic surfactants (e.g.
Dettox) to clean the console
• Failure to carry out recommended routine maintenance as per
these Instructions for Use may reduce the operational lifetime
of the product
• There are no user serviceable parts inside. Contact authorised
service representative for further information
• At product end of life dispose of in accordance with local
environmental guidelines (WEEE)
• In the unlikely event that the probe fails to defrost, immediately
switch off the console and apply saline solution to the probe
• Do not obstruct the vent holes on the disposable probe (located
on the tube) and reusable probe, located as indicated below