13. Annex I – EMC statement and guidelines
Guidance and manufacturer’s declaration – electromagnetic immunity
The Cryomatic is intended for use in the electromagnetic environment specified below. The customer or user should assure that it is used in such
an environment.
Immunity test
IEC 60601 Test level
Compliance level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80MHz to 2.5GHz
3 Vrms
3 V/m
Portable and mobile RF communications equipment should be used no
closer to any part of the Cryomatic, including cables, than the recommended
separation distances calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d = 1.2
√
p
d = 1.2
√
p 80MHz to 800 MHz
d = 2.3
√
p 800MHz to 2.5GHz
Where p is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended
separation distance in metres(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic
site survey
a
, should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
Note 1 At 80MHz and 800MHz, the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
a Field strengths from fixed transmitters, such as base stations (cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the
Cryomatic is used exceeds the applicable RF compliance level above, the Cryomatic should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orientating or relocating the Cryomatic.
b Over the frequency range 150kHz to 80 MHz, field strengths should be less than 3 V/m.