11. Troubleshooting guide
The following table provides a troubleshooting guide for the Cryomatic MKII system error messages.
Error message
Possible cause
Corrective action
F-01 Gas Check Fault
Faulty valves and/or pressure sensors.
Contact distributor or manufacturer for assistance.
F-02 P2 Error
Pressure sensor error.
Contact distributor or manufacturer for assistance.
F-03 P3 Error
Pressure sensor error.
Contact distributor or manufacturer for assistance.
F-04 Probe Blocked
Partial or complete probe blockage.
Re-purge the probe.
Use an alternative probe.
Contact distributor or manufacturer for assistance.
F-05 V1 Error
Solenoid valve error.
Contact distributor or manufacturer for assistance.
F-06 V2 Error
Solenoid valve error.
Contact distributor or manufacturer for assistance.
F-07 V3 Error
Solenoid valve error.
Contact distributor or manufacturer for assistance.
F-008 Over Pressure
Cylinder pressure too high.
Contact distributor or manufacturer for assistance.
Check cylinder pressure does not exceed recommended
maximum pressure (83bar).
Keeler recommends fitting a pressure regulator if cylinder
pressure cannot be maintained below the maximum.
F-09 V1 Error
Solenoid valve leaking.
Contact distributor or manufacturer for assistance.