Instructions for use SURGmatic S201 XL Pro - 1.013.7541 SURGmatic S201 L Pro - 1.013.7540

4 Startup and shut-down | 4.1 Checking the water quantity

12 / 28

Carefully plug the coolant hose in central position onto the media tube pro-
ceeding in axial direction.

Содержание SURGmatic S201 XL Pro

Страница 1: ...SURGmatic S201 L Pro SURGmatic S201 XL Pro Instructions for use SURGmatic S201 XL Pro 1 013 7541 SURGmatic S201 L Pro 1 013 7540 ...

Страница 2: ...istributed by KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Phone 847 550 6800 Fax 847 550 6825 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com ...

Страница 3: ...ions 16 6 2 Troubleshooting 16 6 2 1 Cleaning the spray clip and the spray tube 16 7 Processing steps in accordance with EN ISO 17664 17 7 1 Preparations at the site of use 17 7 2 Non fixing preliminary cleaning of the spray clip and spray tube 17 7 3 Manual processing 18 7 3 1 Manual internal and external cleaning and internal and external disinfection 18 7 3 2 Manual drying 18 7 4 Automated proc...

Страница 4: ...ions please contact KaVo Repair Service KaVo Dental Technologies LLC 11727 Fruehauf Drive Charlotte NC 28273 USA Toll free Direct Customer Service 1 888 ASK KAVO 888 275 5286 Email techservice kavokerr com www kavo com Target group The instructions for use are intended for medical professionals in particular dentists and office personnel The section on Commissioning is also intended for service te...

Страница 5: ...ge conditions Temperature range Transportation and storage conditions Air pressure Transportation and storage conditions Humidity range Protect from moisture Keep dry Fragile handle with care HIBC Code Hazard levels The warning and safety notes in this document must be observed to prevent personal injury and property damage The warning notes are designated as shown below DANGER In cases which if n...

Страница 6: ...Instructions for use SURGmatic S201 XL Pro 1 013 7541 SURGmatic S201 L Pro 1 013 7540 1 User instructions 6 28 NOTICE In cases which if not prevented can lead to property damage ...

Страница 7: ... disposal 2 2 Technical condition A damaged device or components can injure patients users and third parties Use the device and components only if there is no damage on the outside Check to make sure that the device is working properly and is in satisfac tory condition before each use Have parts with sites of breakage or surface changes checked by the Ser vice If the following defects occur stop w...

Страница 8: ...62 may only be combined with the SURGmatic S201 XL Pro 1 013 7541 shank The lack of control equipment for changing the speed range and the direction of rotation can lead to injury Control facility for changing the speed and the direction of rotation must be present The medical device may only be combined with a treatment centre control unit released by KaVo Comply with the Instructions for Use of ...

Страница 9: ...nt and the light for illumination the operation area Surgical burs and cutters with straight handpiece shank or with contra angle shank according to ISO 1797 can be used The handpieces use external spray allowing external irrigation to the working area Based on the speed adjusted in the surgical control unit the handpiece bur rotates up to 40 000 rpm The handpieces interacts with the pa tient thro...

Страница 10: ...01 XL Pro Drive speed max 40 000 rpm max 40 000 rpm Labeling 1 green ring 1 green ring Speed transmission 20 1 20 1 Max torque 55 Ncm 55 Ncm Max torque for the use of hexagon shanks 80 Ncm With push button chuck Usable with surgical burs or diamonds with inner cooling or surgical tools with a hexagon head socket at the shaft Internal cooling system acc to Kirschner and Meyer and external cooling m...

Страница 11: ...ufficient amount of cooling water Thermal damage to the dental pulp Set the water amount for the spray cooling to a minimum of 50 ml min CAUTION Hazard from insufficient amount of spray water Insufficient spray water can cause the medical device to overheat and damage the tooth Check the spray water channels and clean the spray tube with the nozzle needle Mat no 0 410 0931 according to need Switch...

Страница 12: ...Gmatic S201 XL Pro 1 013 7541 SURGmatic S201 L Pro 1 013 7540 4 Startup and shut down 4 1 Checking the water quantity 12 28 Carefully plug the coolant hose in central position onto the media tube pro ceeding in axial direction ...

Страница 13: ... not properly locked in place it can become detached during treatment Do not mount or remove the head while it is rotating Before each treat ment check that the head is firmly seated and that the clamping ring is tight Rotate the clamping ring in the direction of the arrow until it hits the stop and hold it there Insert the head to the stop Make sure that the catches engage properly Rotate the cla...

Страница 14: ...nto place Pull on the medical device to make sure that it is securely affixed to the coupling 5 4 Pulling the contra angle handpiece off the motor coupling Unlock the medical device from the motor coupling by twisting it slightly and then pulling it along its axis 5 5 Inserting the bur or diamond Note Only use carbide cutters or diamond drill bits made by Nobel Biocare that meet EN ISO 1797 type 1...

Страница 15: ... bur to make sure that the chucking system works properly and that the bur is held securely Wear gloves or finger guards when you check insert or remove the insert to prevent injury and infection Insert the bur into the segment of the head drive by twisting the tool slightly and push it to the bur stop Activate the button if necessary Make sure that the tool is seated securely by pulling on it 5 6...

Страница 16: ...s while working Service the medical device When the speed drops or is uneven Service the medical device Missing O ring on the motor coupling Replace O ring See also 2 Instructions for use of motor 6 2 Troubleshooting 6 2 1 Cleaning the spray clip and the spray tube CAUTION Hazard from insufficient amount of spray water Insufficient spray water can cause the medical device to overheat and damage th...

Страница 17: ...d spray tube The non fixing preliminary cleaning is a central constituent and must be per formed prior to the automated processing Accessories required Demineralized water 30 C 2 C 86 F 3 6 F Nozzle needle Brush e g medium hard toothbrush Disposable syringe Note Before cleaning take the hose off the spray clip and the contra angle hand piece Check the patency of the spray clip and spray tube and c...

Страница 18: ...ultra sonic cleaner 7 3 1 Manual internal and external cleaning and internal and external disinfection Not applicable 7 3 2 Manual drying Clean the outside and inside with compressed air until no drops of water are visible Immediately after drying lubricate the KaVo medical device with care agents from the KaVo care system See also 2 7 5 Care products and systems Servicing Page 20 7 4 Automated pr...

Страница 19: ...ing is required 7 4 1 Automated internal and external cleaning and internal and external disinfection Note Prior to cleaning or disinfection in a washer disinfector attach the head to an appropriate base KaVo recommends washer disinfectors in accordance with ISO 15883 1 which are operated using alkaline cleaning agents For program settings as well as cleansers and disinfectants to be used please r...

Страница 20: ...been tested for proper use on our products 7 5 1 Servicing with KaVo Spray Note Head can be serviced alone or attached to a shank Remove the bur Cover the product with the Cleanpac bag Plug the product onto the spray head and press the spray button for one second Servicing the chucking system KaVo recommends servicing the chucking system once weekly Remove the bur place the spray nozzle in the ope...

Страница 21: ... must be mounted there Press the handpiece together with the guide bush of the chuck to be ser viced against the tip of the service coupling Press the button marked with the chuck service symbol Note Close the chuck service mode Option 1 Place the dental handpieces in the QUATTROcare PLUS 2124 A close the front door and start the service procedure Option 2 After three minutes with no service proce...

Страница 22: ...the product so that the packaging is not stretched The quality and use of the sterilization packaging must comply with applica ble standards and be suitable for the sterilization procedure Seal each medical device individually in a sterilization item package 7 7 Sterilization Sterilization in a steam sterilizer autoclave in accordance with ISO 17665 1 CAUTION Improper service and care Risk of inju...

Страница 23: ...ty method at least 10 minutes at 135 C 275 F drying time 30 minutes or alternatively at least 30 minutes at 121 C 250 F Drying time 15 min Use in accordance with the manufacturer s Instructions for Use 7 8 Storage Processed products must be stored protected from bacteria to the extent pos sible and dust in a dry dark cool room Note Comply with the expiration date of the sterilized items 23 28 ...

Страница 24: ...ling piece 0 593 0361 Nozzle pin 0 410 0931 Instrument stand 3 005 5204 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 USA only Material summary Mat No KaVo Spray USA and Canada 2113 A 0 411 9660 QUATTROcare plus Spray USA and Canada 2141 P 1 005 4524 Chuck care set 1 003 1253 Canada only Material summary Mat No KaVo Spray Canada 2114 A 0 411 9680 QUATTROcare plus Spray Canada 214...

Страница 25: ...e liable for defects and their consequences that have arisen or may arise from natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connection instructions calcination or corrosion contaminated air or water supplies or chemical or electrical factors deemed abnormal or impermissible in accordance with KaVo s instructions for use or other manufacturer s...