4 Startup and shut-down | 4.1 Checking the water quantity
4 Startup and shut-down
WARNING
Hazard from non-sterile products.
Infection hazard for dentist and patient.
▶
Before first use and after each use, process and sterilize the medical device
and accessories accordingly.
WARNING
Dispose of the product in the appropriate manner.
Infection hazard.
▶
Before disposal, process and sterilize the product and accessories appropri-
ately.
4.1 Checking the water quantity
CAUTION
Overheating of the tooth due to insufficient amount of cooling water.
Thermal damage to the dental pulp.
▶
Set the water amount for the spray cooling to a minimum of 50 ml/min!
CAUTION
Hazard from insufficient amount of spray water.
Insufficient spray water can cause the medical device to overheat and damage
the tooth.
▶
Check the spray water channels and clean the spray tube with the nozzle
needle (
Mat. no. 0.410.0931
) according to need.
▶
Switch off spray-air and spray-water supply on the treatment device.
▶
Cool the dental bur or diamond via the external and/or internal feed, if
available. The supply can take place individually or via a coupling piece.
▶
During surgical interventions, comply with the necessary precautions re-
garding cooling.
▶
Use physiological, sterile cooling fluid.
▶
Ensure that the coolant supply is free of air.