6 Checking for malfunctions and troubleshooting
6.1 Check for malfunctions
CAUTION
Product heats up.
Burn injury or product damage due to over-heating.
▶
Do not continue working if the product heats up irregularly.
▶
The medical device overheats while working:
Service the medical device.
▶
When the speed drops or is uneven:
Service the medical device.
▶
Missing O-ring on the motor coupling:
Replace O-ring.
See also:
2
Instructions for use of motor
6.2 Troubleshooting
6.2.1 Cleaning the spray clip and the spray tube
CAUTION
Hazard from insufficient amount of spray water.
Insufficient spray water can cause the medical device to overheat and damage
the tooth.
▶
Check spray water channels and if necessary clean spray tubes with the
nozzle needle (
Mat. no. 0.410.0931
).
▶
Use the nozzle pin (
Mat. no. 0.410.0931
) to free the water passage on
the spray clip on both sides.