Operating Instruction STERIclave B 2200 / 2200 P
6 Operation
|
6.10 Changing the water
6.10.2 Filling the fresh water tank
Requirement
The H
2
O min. LED will light up if there is not enough water in the fresh water tank.
The water level will need to be topped up after approx. 5 cycles.
See also: 7.4 Regular total cleaning, Page 61
▶ Fill the fresh water tank with high-quality distilled or demineralised water (con‐
ductivity ≤ 15 μS/cm; hardness ≤ 0.02 mmol/l).
See also: 4.4.1 Fill the fresh water tank, Page 29