Operating Instruction STERIclave B 2200 / 2200 P
6 Operation
|
6.4 Selecting and starting a program
6.4 Selecting and starting a program
6.4.1 Overview of programs
Display
Program
Total cycle test
(Including drying)
Parameter
Load
(test according to DIN EN
13060:2004)
Type
(DIN EN 13060:2004)
LED 1
Sterilise 1
18 min
134 ℃
5 min
3 pre-vacuum pha‐
ses
Drying 5 min (3
min vacuum, 1 min
ventilation)
all sterilisation
items up to 134℃
for example: In‐
str turbi‐
nes
(Helix test)
B
LED 2
Sterilise 2
36 min
121℃ 20 min
3 pre-vacuum pha‐
ses
Drying 11 min (4
min vacuum, 2 min
ventilation)
Solid sterilisation
items up to 121℃
un single
packed
(Helix test)
B
LED 3
Fast cycle
15 min
134℃ 4 min
2 pre-vacuum pha‐
ses
drying 3 min
(2 min vacuum, 1
min ventilation)
Solid sterilisation
items up to 134℃
unpacked
S
LED 4
SPECIAL
S1
Prions
32 min
134 ℃
19 min
3 pre-vacuum pha‐
ses
Drying: 3 + 1 min
Against Creutz‐
feld-Jacob and
BSE pathogens*
B
SPECIAL
P2
Maximum load
20 min
134 ℃
5 min
4 pre-vacuum pha‐
ses
Drying: 3 + 1 min
Difficult materials
such as textiles
(Helix test)
B
SPECIAL
P3
Maximum load
38 min
121 ℃
20 min
4 pre-vacuum pha‐
ses
Drying: 4 + 2 min
Difficult sterilisati‐
on items
Up to 121°C
(for example: sur‐
gical cassettes)
B
SPECIAL
P4
Individual
15 to 80 min
Programmable by
the user:
Temperature:
121-135 ℃
Time: 3-60 min.
Pre-vacuum pha‐
ses: 2.3 or 4
Drying with pre-va‐
cuum and ventila‐
tion: 2+1, 4+2 or
6+3 min
Depending on the
set values
45/78