Prepare the unit
▶ Switch the power switch to "Off".
▶ Place the SMARTair plus mobile in a horizontal position.
▶ Slightly pull apart the grip strips ② to unlock the snap locks.
▶ Remove the cover ① of the SMARTair plus mobile.
Exchanging the filter bag
▶ Carefully remove the holder ① from the filter connector ⑤, and remove the filter.
▶ Seal the filled filter bag with the adhesive tab ② and throw it away ③.
▶ Check if the filter compartment contains dust and clean if necessary.
▶ Shove the new filter bag with the cardboard frame ⑥ onto the filter connector
⑦, fix it with the holder ④, and check to see if it is firmly seated.
▶ Remount the cover and lock in place.
▶ Move the SMARTair plus mobile back into operating position.
7.2.3
Changing the adsorption filter of the SMARTair plus
Note
The adsorption filter must be changed every six months.
Instructions for use SMARTair plus, SMARTair plus mobile
7 Cleaning and servicing | 7.2 Servicing
39/57