To switch the positioning lights on (if they are not already on), do one of the
following:
a.) Press the thumb wheel on the underside of the patient support table.
b.) Press any of the positioning controls (button or joystick).
The lights automatically switch off after two minutes. To switch them off
earlier, press the positioning joystick.
8.7.1 Moving image volume vertically (Z laser)
About this task
The volume bottom light (Z laser) indicates the position where the lower
edge of the image volume is.
Steps
1. Check that the image volume is positioned at the correct height for your
patient.
If you need to move the image volume up or down, press the image
volume up (1) or down (2) button. The volume bottom light (Z laser)
moves accordingly on your patient’s face.
8 3D patient exposure
Instructions for use
KaVo ProXam 43