![KaVo ProXam 3DQ Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/kavo/proxam-3dq/proxam-3dq_instructions-for-use-manual_3954372044.webp)
The exposure values have been preset at the factory for each patient size,
image resolution and ULD (Ultra Low Dose) setting. The preset exposure
values are average values and they are only meant to guide the user.
The preset exposure values are shown in the following tables.
Factory presets for image resolution Low dose in programs Nose and Sinus
Patient size
kV value
mA value
mA value with
ULD
Child (XS)
90
2.8
1.4
Small adult (S)
90
3.6
1.8
Medium-sized adult (M)
90
5.0
2.2
Large adult (L)
90
5.6
2.8
Extra large adult (XL)
90
7.1
3.6
Factory presets for image resolution Low dose in other programs
Patient size
kV value
mA value
mA value with
ULD
Child (XS)
90
4
2
Small adult (S)
90
5
2.5
Medium-sized adult (M)
90
6.3
3.2
Large adult (L)
90
8
4
Extra large adult (XL)
90
10
5
Factory presets for image resolution Normal in programs Nose and Sinus
Patient size
kV value
mA value
mA value with
ULD
Child (XS)
90
3.2
2.2
Small adult (S)
90
4
2.8
Medium-sized adult (M)
90
5
3.6
Large adult (L)
90
6.3
4.5
Extra large adult (XL)
90
8
5.6
Factory presets for image resolution Normal in other programs
Patient size
kV value
mA value
mA value with
ULD
Child (XS)
90
5
3.6
Small adult (S)
90
6.3
4.5
Medium-sized adult (M)
90
8
5.6
Large adult (L)
90
10
7.1
Extra large adult (XL)
90
11
9
8 3D patient exposure
40 KaVo ProXam
Instructions for use
Содержание ProXam 3DQ
Страница 1: ...KaVo ProXam 3DQ Instructions for use 3D imaging EN...
Страница 8: ...General warning Standard ISO 7010 3 Symbols on product labels 4 KaVo ProXam Instructions for use...
Страница 78: ...Patient handles and touch screen 12 Cleaning and disinfection 74 KaVo ProXam Instructions for use...
Страница 87: ......
Страница 88: ...KaVo Dental GmbH Bismarckring 39 88400 Biberach Deutschland www kavo com 1 015 1242...