4 Installation | 4.3 Assembly of dentist chair PHYSIO Evo F
19 / 40
4.2.4 Adjust the gas-pressure spring
NOTE
The pressure point of the gas-pressure spring is set at the factory. If the seat
of the dentist chair PHYSIO Evo/EvoF/One is not adjusted securely, the pres-
sure point of the gas-pressure spring must be adapted.
Also refer to:
6.1.1 Adjust the seat height, Page 29
4
Incline the seat forwards and remove the upholstery.
4
Adjust the pressure point as described below using a Phillips screwdriver:
screw in screw, the gas-pressure spring is released earlier.
unscrew screw, the gas-pressure spring is released later.
4
Remount the seat upholstery.
4.3 Assembly of dentist chair PHYSIO Evo F
NOTE
The chair is delivered with soft castors for hard flooring as standard.
Use suitable castors for the relevant flooring.
Only use soft castors for hard flooring. Only use hard castors for soft flooring.