3-5
Software Backup and Restore, Recovery, and Updates
032-0358-EN Rev B
Restore Operating Environment
The operating system image on the scanner controller (model 800137) is pre-loaded at the factory. Sites
receive a USB Flash Drive (UFD) of this image. Follow the procedures below in case the operating
environment must be restored.
Boot the Scanner Controller from the Bootable UFD
1.
Insert the bootable UFD into an available USB port on the scanner controller.
2.
Boot the scanner controller (power off/power on) to the BIOS setup screen by pressing the
Delete
key at the beginning of the boot sequence, typically when the SmartScan STUDIO splash screen
is displayed and beeps are audible. You may want to press
Delete
several time when the splash
screen is displayed to ensure entry into the BIOS setup screen.
3.
Press the
Right Arrow
key until the
BOOT
options tab is selected, and then press the
Up Arrow
key to select the
Hard Drive BBS Priorities
option.
4.
Press the
Enter
key twice to access the hard drive boot options. Use the
Down Arrow
key to select
the UFD entry (indicated by the device description of the UFD).
5.
Press the
Enter
key to select the UFD.
6.
Press the
F4
key to display a save dialog, then select
Yes
and press the
Enter
key to save the
change and initiate a reboot. The scanner controller will reboot from the UFD.
7.
From this point, continue with one of the following operations:
•
Install the Operating System Image (page
3-5)
•
Make a Backup Copy of the Software Package onto a Blank UFD (page
3-9)
Install the Operating System Image
NOTE:
The touch screen interface is not active in this application. Attach a mouse or use keyboard
commands (tab to move between fields, up and down arrow keys to make selections in a field,
space bar or Enter key to enter).
1.
Obtain the bootable UFD and follow instructions in
Boot the Scanner Controller from the Bootable
UFD (page
3-5)
.
2.
After the scanner controller has booted from the
UFD, the Installation Progress Monitor is displayed,
followed by a window with the message:
Verifying
packages
. Wait (about five minutes) until the
message
Package verification passed
is
displayed and options on the window become
active. If package verification fails, obtain a new
bootable UFD.
3.
Select
Install
from the Select Action drop-down
menu.
Содержание OP 3D Vision
Страница 1: ...Service Manual Cone Beam 3D 2D Panoramic Dental Imaging System KaVo OP 3D Vision...
Страница 22: ...2 16 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B...
Страница 48: ...6 8 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B...
Страница 50: ...7 2 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B Schematic Sheet 1 of 3...
Страница 51: ...7 3 System Schematics 032 0358 EN Rev B Schematic Sheet 2 of 3...
Страница 52: ...7 4 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B Schematic Sheet 3 of 3...
Страница 53: ......