3-7

Software Backup and Restore, Recovery, and Updates

032-0358-EN  Rev B

6.

Select 

Upgrade

 from the Select Action drop-down menu.

7.

Select 

Start

. A warning is displayed that the inactive partition on the PC will be erased.

8.

Select 

Continue

 when prompted. The software update continues and status is displayed.

9.

Review status messages as needed. When the Quit button becomes active, select 

Quit

.

 

10. Select 

 to reboot and complete update. The scanner controller will reboot and start first-time 

boot processing (5-10 minutes). It will reboot again, and perform checks and configuration, then 
display the Login screen. No manual system configuration is required since the previous 
configuration has been maintained.

11. Remove the UFD from the USB port.

12. A unique software license is now required to establish the accessible field of view on the V-series 

upgradeable scanners. A license file is included on the installer media that should contain the 
unique license. If the license is missing, you will not be able to access the protocol screen in 
SmartScan STUDIO. 

a.

Log in, select 

Acquire

 from the menu, and navigate to the Protocol screen. 

b.

Check that the screen is accessible and that the protocols displayed are for the correct 
system. Refer to the appropriate 

User Manual

 for a list of protocols for each system. 

c.

Contact Technical Support if a license is required.

13. Perform updates to clinical software installed at the site. See 

Update Clinical Software (page

 

3-7)

.

Update Clinical Software

Clinical software at the site must be updated to the most current versions after the scanner controller 
software is updated. Access Control Panel at each site workstation to view the current clinical software 
versions loaded. 

CAUTION

CAUTION

It is REQUIRED that the most current versions of the clinical software are installed after the 
scanner controller is updated. The software checks for mismatched versions. A database error will 
be indicated on the SmartScan STUDIO Manager status display if versions do not match.

All servers and client workstations must be running one of the following operating systems. The update 

software performs a check, and will not continue if the computer is not running one of these operating 
systems.

Windows

®

 7 Professional, Ultimate, and Enterprise (64-bit) SP1

Windows

®

 8.1 Pro and Enterprise (64-bit)

Windows

®

 10 Pro and Enterprise (64-bit)

Windows

®

 Server 2008 R2 SP1

Windows

®

 Server 2012 (all versions)

Windows

®

 Server 2012 R2 (all versions) 

Windows

®

 Server 2016 (all versions) 

Содержание OP 3D Vision

Страница 1: ...Service Manual Cone Beam 3D 2D Panoramic Dental Imaging System KaVo OP 3D Vision...

Страница 2: ...ves or others to unnecessary radiation It is important that you become fully acquainted with applicable government radiation protection regulations Many provisions of these regulations are based on th...

Страница 3: ...hair Assembly Parts List 2 13 Upper Chair Assembly 2 14 Upper Chair Assembly Parts List 2 15 Chapter 3 Software Backup and Restore Recovery and Updates Backup and Restore 3 1 Backup Files 3 1 Restore...

Страница 4: ...nd Codes Home Commands 5 1 Movement Commands 5 1 Read Commands 5 2 Set Commands 5 4 X Ray Commands 5 5 Miscellaneous Commands 5 5 Chapter 6 Safety Checks Recommended Safety Checks 6 1 Safety Tests 6 1...

Страница 5: ...h the plastic cap only 1 Press the OFF button on the operator control box The scanner shuts down and the POWER indicators on the operator control box and scanner go OFF 2 At rear of overhead switch ci...

Страница 6: ...r panel fits very tightly over the panel If this panel cover is removed and re installed for any reason perform the following Crosshair Laser Adjustment Chair Calibration Centerline Laser Adjustment G...

Страница 7: ...ny covers open or cables removed To ensure power to the scanner is locked out from the main supply 1 At rear of scanner turn circuit breaker off 2 Remove the overhead cord bracket 3 Unplug the scanner...

Страница 8: ...2 2 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B System Assemblies 1 2 5 7 12 7 13 4 11 6 3 8 8 9 10...

Страница 9: ...3 8 Long 4 1 013 0489 Resident Indicator Panel 5 1 011 1964 Receptor Assembly 6 1 013 1578 Chair Assembly 7 1 011 4086 Leveler 3 8 16 LG HX BC 1000 Cap 8 0 830 0200 Leveler 3 8 16 LG BC 1000 Cap 9 1 0...

Страница 10: ...2 4 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B Overhead Assembly 12 1 2 3 4 5 6 7 8 9 13 14 15 17 15 16 15 18 19 20 21 10 11 22 23 24 25 26 27 28 29 30...

Страница 11: ...ansformer 13 SHCS 5 16 18 x 1 Long 14 1 010 0953 Power Bracket Assembly 15 1 007 4094 Buss HTB 84M 16 1 006 3075 10A Fuse Digi Key F985 ND 17 1 006 3074 2 5A Fuse Digi Key F979 ND 18 1 006 3076 0 25 A...

Страница 12: ...2 6 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B Rotation Drive Assembly 1 2 3 4 5 6 7 8 9 10 11 12 14 13 15...

Страница 13: ...10 3 1 007 4102 Drive Gear 4 1 006 8180 Gear Encoder Disk 5 1 007 4105 Gear Clamp 6 1 007 4106 Rotation Motor Bracket 7 1 007 4107 Al Round Spacer 8 1 007 3996 Gear Encoder 9 Split Washer 6 10 SHCS 6...

Страница 14: ...2 8 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B X Ray Source Assembly 1 3 4 5 6 7 8 9 11 10 13 14 15 16 17 18 12 2 19...

Страница 15: ...ly 6 1 005 8622 Cross Laser Assembly 7 0 830 0492 Cross Laser Mirror 8 BHCS 6 32 x 1 4 Long 9 Split Washer 10 10 SHCS 10 24 x 3 8 Long 11 1 010 9900 Beam Limiter 12 1 013 0493 Patient Align Panel 13 0...

Страница 16: ...2 10 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B Flip Receptor Panel Assembly 1 Panel Side Panel 2 4 3 5 6 11 10 9 8 12 14 15 7 16 18 17 19 Drive Side 20 13...

Страница 17: ...5 Rear Receptor Cover Assy 6 1 007 4114 Stop Screw 7 1 007 4001 Drive Screw Assembly 8 SHCS M3 x 8mm Long 9 1 005 8630 Flip Stepper Motor 10 1 007 4003 Idler 11 1 007 4115 Cable Shield 12 Split Washer...

Страница 18: ...2 12 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B Chair Assembly 1 5 2 4 3 8 7 6 9 10 11 12 13 14 17 18 20 19 See Upper Chair Assembly 15 16...

Страница 19: ...r Junction Cable 5 SHCS 1 4 20 x 5 8 Long 6 1 4 Split Washer 7 1 007 4122 Rail 8 0 830 2024 Slider Plate 9 1 007 4125 Wheel 10 1 007 4126 Wheel Axle 11 1 007 4127 Actuator Plate 12 1 007 4128 Receptac...

Страница 20: ...14 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B Upper Chair Assembly 1 10 2 3 4 6 7 8 9 5 11 12 13 14 15 16 17 18 20 19 21 22 25 26 27 24 23 28 30 31 32 29 33 34 35 36 37 43 42 41 40 39 38 44 4...

Страница 21: ...r 10 18 SHCS 10 32 x 5 8 Long 19 SHCS 1 4 20 x 1 1 4 Long 20 Split Washer 1 4 21 Flat Washer 1 4 22 0 830 4997 Hinge Bar 23 1 007 4140 Bunting BM4971 24 1 007 4010 Tee Nut 25 1 007 4141 Magnet Plate 2...

Страница 22: ...2 16 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B...

Страница 23: ...ed by the site need to be manually copied and restored This data is not stored in the two files above The recommended workflow is to use a USB connected device USB Flash Drive or external drive with a...

Страница 24: ...Select drive location of the USB device d To save backup files to a specific folder browse to desired folder location Otherwise a folder will be created on the media with the backup job name e Select...

Страница 25: ...bottom of the window NOTE The duration of the backup operation will vary depending on the amount of data being backed up 8 When backup job completes Last backup status field will indicate status NOTE...

Страница 26: ...store operation will vary depending on the amount of data being restored 7 A progress bar is displayed at the bottom of the window When complete select View Job Logs Last Restore Log view a log of the...

Страница 27: ...row key to select the UFD entry indicated by the device description of the UFD 5 Press the Enter key to select the UFD 6 Press the F4 key to display a save dialog then select Yes and press the Enter k...

Страница 28: ...tep b b Obtain the Imagers DVD for the panel Obtain an external DVD drive or copy the DVD contents to a UFD c Navigate to the Imagers content on the DVD or UFD d Start Setup exe and follow instruction...

Страница 29: ...he screen is accessible and that the protocols displayed are for the correct system Refer to the appropriate User Manual for a list of protocols for each system c Contact Technical Support if a licens...

Страница 30: ...missions set for the SmartScan STUDIO Integration Services services are deleted during an update and will need to be reset Copy a License File for V Series Upgradeable Systems 1 If license file is on...

Страница 31: ...is not corrected or the error persists Make a Backup Copy of the Software Package onto a Blank UFD NOTE The touch screen interface is not active in this application Attach a mouse or use keyboard comm...

Страница 32: ...tating no update was performed 14 Select The scanner controller will reboot to the Login screen SmartScan STUDIO Manager Settings Window In addition to providing options for configuring SmartScan STUD...

Страница 33: ...ote Helpdesk assistance Backup Restore Access FBackup for backing up or restoring files on the scanner controller Refer to Software Backup Recovery and Updates Explorer Displays Windows Explorer Contr...

Страница 34: ...ive values shift the collimation out negative values shift the collimation in 6 Press to initiate a scout scan Review scout scan for evidence of beam limiter shutter collimation on all four borders NO...

Страница 35: ...ubsequent MR 100000 commands will move back to this position HSR Home Rotation Short Performs a quick check within approximately 1 degree of the current location for home If home is not located the ro...

Страница 36: ...are unique in that they occur at any time when their respective motors are not in the home position which occurs practically any time the machine is in use For example if a stall occurs it will do so...

Страница 37: ...oller error 6 0b 0000 0000 0000 0000 0010 0000 32 Encoder direction 7 0b 0000 0000 0000 0000 0100 0000 64 Rotation optical switch status 8 0b 0000 0000 0000 0000 1000 0000 128 Rotation limit switch 9...

Страница 38: ...range setting Command syntax RX n 0 120 KV Range default 1 90 KV Range Results returned in format OK kV mA TicksPerFrame XrayOn XrayOff PanelRead PulseMode Set Commands SP Sets platform position Sets...

Страница 39: ...and watchdog settings Command syntax XCT Distance Speed Watchdog Distance 1000 to 668563 Speed 1000 to 65000 Watchdog 0 disabled default 1 enabled XIP Performs a Panoramic scan for the requested param...

Страница 40: ...format OK mv x hw x sw x bv x sn x mv Machine Version hw Hardware Version sw Software Version bv Beam Limiter Version sn Machine Serial Number RSN Read Serial Number Command syntax RSN Reads the curre...

Страница 41: ...ogies standards and local regulations The personnel assessing the safety must be able to recognize possible consequences and risks arising from non conforming equipment Responsibility is accepted for...

Страница 42: ...temperature humidity atmospheric pressure and mains voltage present at the test site Visual Inspection Perform a visual inspection of system and accessories to include Safety related markings and labe...

Страница 43: ...of scanner power cord and push test button on probe 6 Check reading on meter Acceptable reading is a value less than 3 ohms 7 Leave meter in place receptor cover off and scanner unplugged for Insulat...

Страница 44: ...black clamp on grounded metal part WARNING The next step will apply 500 VDC to probe Do not touch probe while test button on probe is engaged 5 Place red probe across the line and neutral prongs of sc...

Страница 45: ...7 Remove meter and replace overhead cover Functional Test This test is used to ensure proper functioning of the equipment indicators mechanical systems and safety devices Perform an interrupted PAN sc...

Страница 46: ...the patient e stop X rays are NOT emitted during a dry run a Seat a person in the scanner with the patient e stop b In SmartScan STUDIO select Same Exam from dialog c Select any CT protocol and press...

Страница 47: ...e Passes Ground Check Earth Leakage Current All readings equal to or below limit Insulation Resistance Functional Test Passes equipment operational Deficiency Note Overall assessment No safety or func...

Страница 48: ...6 8 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B...

Страница 49: ...tem schematics Ensure power is removed before servicing the scanner WARNING High voltage is present in the scanner Remove power from scanner before removing covers or cables To avoid personal injury f...

Страница 50: ...7 2 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B Schematic Sheet 1 of 3...

Страница 51: ...7 3 System Schematics 032 0358 EN Rev B Schematic Sheet 2 of 3...

Страница 52: ...7 4 KaVo OP 3D Vision Service Manual 032 0358 EN Rev B Schematic Sheet 3 of 3...

Страница 53: ......

Страница 54: ...C 1910 North Penn Road Hatfield PA 19440 USA Tel 1 215 997 5666 Fax 1 215 997 5665 032 0358 EN Rev B 2017 October 1 Kaltenbach Voigt GmbH Bismarckring 39 88400 Biberach Germany Tel 49 7351 56 0 Fax 49...

Отзывы: