Care instructions

Primus 1058 Life

Содержание KaVo Primus 1058 Life TM

Страница 1: ...Care instructions Primus 1058 Life...

Страница 2: ...Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488...

Страница 3: ...g the instrument holder 20 2 5 2 Preparing the connector for the suction hose 20 2 5 3 Reprocessing of sieve inserts and suction hoses 21 2 6 Preparation of the patient unit 21 2 6 1 Preparation of th...

Страница 4: ...bly instructions TI Technician s instructions SC Safety checks IEC International Electrotechnical Commission RI Repair instructions EMC Electromagnetic compatibility Scaler Scaler ZEG IR Infrared IrDA...

Страница 5: ...e device off NOTICE Product damage due to improper disinfection Malfunctions Use the disinfectant according to the specifications of the manufacturer No spray disinfection perform wipe disinfection on...

Страница 6: ...disher Z For programme settings as well as cleansers and disinfectants to be used please refer to the Instructions for Use of the washer disinfector Directly after automated cleaning disinfection trea...

Страница 7: ...sation All sterilisable products are temperature resistant up to max 138 C Select a suitable procedure depending on the available autoclave from the following sterilisation processes Autoclave with th...

Страница 8: ...for cleaning Adjust the backrest to vertical position and move the chair upward Also refer to Instructions for usePrimus 1058 Life Turn device off Switch main device switch off Cleaning and disinfecti...

Страница 9: ...use of water Damage to the foot control For cleaning the foot control must not be immersed in water or placed un der running water When wiping it off with a moist cloth do not touch the contacts of th...

Страница 10: ...4 Cleaning and disinfection of S table handle and holder pad anti slip mat Press the push button and pull off the handle Clean the handle under flowing water Disinfect the white and grey handle by wip...

Страница 11: ...andard handle Sterilise the holder pad anti slip mat The grey handle Mat no 1 002 4489 can be sterilised see symbol on the handle 2 4 2 Reprocessing of the instrument holder of the TM table Note The h...

Страница 12: ...achment during the cleaning and germ reduction of the instrument holder with patient part assembly kit without patient part assembly kit Pull the latches evenly outward and carefully remove the handpi...

Страница 13: ...isinfection of the instrument arm S table NOTICE Damage caused by treatment in a washer disinfector Do not sterilise the instrument arm S table in a washer disinfector Note Neither the rinsing program...

Страница 14: ...unction and multifunction handpiece Note Wet the O rings of the gripping sleeves after sterilisation using silicone grease Mat no 1 000 6403 or KaVo Rota Spray 2 Mat no 0 411 7510 After replacing the...

Страница 15: ...0o C 5o C Brush e g medium hard toothbrush Disassemble the gripping sleeve and cannula Detach the key part from the gripping sleeve Clean the key part gripping sleeve and cannula by brushing them unde...

Страница 16: ...nula in a disinfectant or ultra sonic bath Drying the 3 function and multifunction handpiece Manual drying Blow off the outside and inside with compressed air until water drops are no longer visible A...

Страница 17: ...sure 2 13 bar NOTICE Damage caused by over heating Damage to the gripping sleeve Immediately remove the parts from the steriliser after the sterilisation cycle is completed Storage Reprocessed product...

Страница 18: ...r Mat no 0 200 3098 and clean the filter holder Note The turbine return air filter is in the same position in the dentist element vari ant with S table 2 4 7 Reprocessing KaVo ERGOcam One Note The cam...

Страница 19: ...age to the camera due to improper disinfection Irreversible destruction of the camera Disinfect with wipes only do not use disinfectant sprays Do not immerse the camera in liquids Do not sterilise the...

Страница 20: ...t unit 2 5 1 Preparing the instrument holder The instrument holder on the assistant unit is prepared in the same way as the instrument holder on the dentist unit See also 2 2 4 2 Reprocessing of the i...

Страница 21: ...ed from the holder Hold the cloth under the sieve seat and pull the sieve insert out of the housing If needed insert a new sieve insert Mat no 07633423 into the seat Wipe the suction tubes with disinf...

Страница 22: ...tion damage Remove the tumbler filler Mat no 0 764 8541 by pulling upwards Take off the tumbler holder Mat no 07646153 Clean the tumbler holder and tumbler filler under running water Wet the O ring Ma...

Страница 23: ...aused by inappropriate cleaning agents Do not use acid containing sand containing and form forming cleaning agents Add a dose of DEKASEPTOL gel to the spittoon bowl and distribute the agent with a bru...

Страница 24: ...th the enclosed dosing attachment Mat no 1 002 0287 add the proper amount of KaVo OXYGENAL 6 Mat no 0 489 3451 with each filling For the correct amount please refer to the instructions of the dosing a...

Страница 25: ...erm reduction of the water supplying system This counteracts the formation of mi cro organisms at times of stagnation and ensures a hygienic water supply for the patients 2 8 1 Rinsing programme The r...

Страница 26: ...e triple function and multifunctional handpieces must be re moved for the rinsing program and the intensive germ reduction S table Lock the swing arm into the front catching position to ensure that th...

Страница 27: ...germ reduction attachment is available for the COM FORTbase For the rinsing programme and intensive germ reduction press the seat ad apter into one of the four rubber seats of the germ reduction atta...

Страница 28: ...nsive germ reduction button to immediately start the process The selected rinse program starts or Turn off the device and do not restart it until the next morning The cycle is started automatically Th...

Страница 29: ...ed The bottom LED of the HYDROclean button flashes Intensive germ reduction must be carried out At first start up Before during and after prolonged downtimes more than 4 weeks Over the weekend During...

Страница 30: ...sive germ reduction can be interrupted in pro gram step 4 e g on the last workday before the weekend Program step 4 is the dwell time in which the microorganisms are killed by an increased OXY GENAL 6...

Страница 31: ...es getting kinked remove hoses from the germ reduction attachment and place them in the dentist support Keep the germ reduction attachment in the spittoon bowl Cover the treatment unit during extended...

Страница 32: ...Gel Aspirate one tumbler of cold water with each suction hose With the suction handpiece attached or by hand press down the DEKA SEPTOL dosing facility once each for the saliva ejector hose and the s...

Страница 33: ...the HYDROclean button This starts up the HYDROclean function The HYDROclean function is completed when the top LED goes dark The process can be discontinued at any time by pressing the HYDROclean but...

Страница 34: ...Do not use spray disinfection Only disinfect by wiping Clean the surfaces with a soft cloth and water and wipe disinfect them with an approved surface disinfectant Approved disinfectants CaviCide Cavi...

Страница 35: ...erties Push the buttons on the bottom of the handle and pull the handle off in the direction of the arrow 1 1 Clean the surfaces with a soft cloth and water and wipe disinfect them with an approved su...

Страница 36: ...uctions When the OXYGENAL 6 container is empty you hear an acoustic signal every 10 seconds Note Replenish KaVo OXYGENAL 6 only if there is a beep every 10 seconds The signal can be turned off as foll...

Страница 37: ...for damage 3 1 1 Replenish OXYGENAL 6 when using the water bottle NOTICE Damage caused by defective water bottle Malfunctions or failures from a leaky water bottle Do not wash the water bottle in a di...

Страница 38: ...Remove the protective hose from the tube of the KaVo OXYGENAL 6 dosing facility Guide the tube through the orifice of the coarse sieve of the KaVo OXY GENAL 6 bottle Screw the KaVo OXYGENAL 6 dosing f...

Страница 39: ...stances remaining in the dosing head are not contamination but rather agents that are deposited there during extended periods of non use after condensation of the water These deposits redissolve after...

Страница 40: ...nfectant 3 2 Check the OXYGENAL 6 concentration 40 44 3 2 Check the OXYGENAL 6 concentration Test the standard concentration 0 02 H2O2 200ppm Fill a measuring cup 40 ml with 39 ml water and 1 ml treat...

Страница 41: ...oncentrations can be tested using the following table H2O2 concentration Measuring cup Ratio water treated water Value shown on test strip 0 02 200 ppm standard concentration 40 ml 39 1 5 mg l 0 05 50...

Страница 42: ......

Страница 43: ......