K4 Table top controller 4954
9
C
• Motor speed too high for longer than
2 sec, possibly one phase of the motor is
interrupted or motor electronics are faulty
R
• Repair fault
C
• Speed preselection potentiometer dis-
connection (open circuit)
R
• Repair fault
C
• No foot switch connected (possibly open
circuit)
R
• Connect foot switch and check function
C
• Foot switch at start when mains switched
on
R
• Release foot switch (stop)
C
• Short-circuit in direct-current chopper
(transistor)
R
• Repair motor electronics
C
• Direct-current chopper or measuring
device U_wr, I_wr faulty
R
• Repair motor electronics
C
• Inverter faulty (automatic test on swit-
ching on)
R
• Repair motor electronics
C
• Brake device faulty (automatic test on
switching on)
R
• Repair motor electronics
C
• Warning - Data loss in EEPROM memory
R
• Start motor, if necessary send to EWL
for checking and recalibration
C
• Program memory (ROM) faulty (auto-
matic test on switching on)
R
• Send to EWL for checking and soldering
in new processor
F
= Fault
C
= Cause
R
= Rectification
F
• flash 20 times
F
• flash 19 times
F
• flash 16 times
F
• flash 15 times
F
• flash 14 times
F
• flash 13 times
F
• flash 12 times
F
• flash 11 times
F
• flash 9 times
F
• flash 6 times